光学相干断层扫描血管造影与免散瞳视网膜相机在评估原发性开角型青光眼视神经乳头参数及与视乳头周围血管密度关系方面的一致性。

The agreement between optical coherence tomography angiography and non-mydriatic retinal camera in estimating the optic nerve head parameters and relations with the peripapillary vessel density in primary open-angle glaucoma.

作者信息

Koylu Mehmet Talay, Yilmaz Hayati, Yilmaz Alper Can

机构信息

Department of Ophthalmology, Gulhane Medical Faculty, Health Sciences University, Ankara, Turkey.

Department of Ophthalmology, Umraniye Training and Research Hospital, Health Sciences University, Istanbul, Turkey.

出版信息

Eye (Lond). 2021 Mar;35(3):959-965. doi: 10.1038/s41433-020-1006-x. Epub 2020 Jun 4.

Abstract

BACKGROUND/OBJECTIVES: To investigate the agreement between optical coherence tomography angiography (OCT-A) and non-mydriatic retinal camera (NMRC) in estimating the optic nerve head (ONH) parameters and evaluate the associations between peripapillary vessel density (VD) with ONH parameters assessed by both devices.

METHODS

The OCT-A and NMRC images of 262 participants (138 eyes were diagnosed with POAG, 63 were glaucoma suspect, and 61 were healthy subjects) were screened retrospectively. The vertical cup-to-disc ratio (VCDR), cup-to-disc area ratio (CDAR), rim area (RA), disc area (DA), and cup volume (CV) were recorded using both devices. Peripapillary VD data were also recorded using OCT-A. The intraclass correlation coefficient (ICC) and Bland-Altman plots were obtained to investigate the agreement between the devices. Age- and intraocular pressure (IOP)-corrected associations between the VD and ONH parameters measured by the two devices were analyzed using linear regression models.

RESULTS

ICC revealed excellent agreement for VCDR and CDAR, but poor agreement for RA, DA, and CV (ICC = 0.801, 0.793, 0.445, 0.298, and 0.527, respectively). The Bland-Altman plots showed poor agreement for VCDR and CDAR with a large span of limits of agreement (0.500 and 0.528, respectively) and significant proportional biases (P < 0.05). VD was found to be strongly associated with the VCDR and CDAR measurements of both devices (P < 0.001), but the associations were stronger for the measurements of NMRC.

CONCLUSIONS

The disagreement between the devices should be considered in clinical practice. The associations between the VD and ONH parameters proved once again the importance of the ONH parameters in terms of the management of glaucoma patients.

摘要

背景/目的:研究光学相干断层扫描血管造影(OCT-A)与免散瞳视网膜相机(NMRC)在评估视神经乳头(ONH)参数方面的一致性,并评估两种设备所测视乳头周围血管密度(VD)与ONH参数之间的相关性。

方法

回顾性筛选262名参与者的OCT-A和NMRC图像(138只眼被诊断为原发性开角型青光眼,63只为青光眼可疑患者,61只为健康受试者)。使用两种设备记录垂直杯盘比(VCDR)、杯盘面积比(CDAR)、视盘边缘面积(RA)、视盘面积(DA)和杯体积(CV)。还使用OCT-A记录视乳头周围VD数据。获得组内相关系数(ICC)和Bland-Altman图以研究两种设备之间的一致性。使用线性回归模型分析两种设备测量的VD与ONH参数之间经年龄和眼压(IOP)校正后的相关性。

结果

ICC显示VCDR和CDAR具有良好的一致性,但RA、DA和CV的一致性较差(ICC分别为0.801、0.793、0.445、0.298和0.527)。Bland-Altman图显示VCDR和CDAR的一致性较差,一致性界限跨度较大(分别为0.50 和0.528)且存在显著的比例偏差(P<0.05)。发现VD与两种设备测量的VCDR和CDAR均密切相关(P<0.001),但NMRC测量的相关性更强。

结论

在临床实践中应考虑两种设备之间的差异。VD与ONH参数之间的相关性再次证明了ONH参数在青光眼患者管理中的重要性。

相似文献

本文引用的文献

3
Optical coherence tomography angiography in glaucoma.光学相干断层扫描血管造影术在青光眼中的应用。
Curr Opin Ophthalmol. 2019 Mar;30(2):110-116. doi: 10.1097/ICU.0000000000000554.
4
Bland-Altman analysis: A paradigm to understand correlation and agreement.布兰德-奥特曼分析:一种理解相关性和一致性的范式。
Turk J Emerg Med. 2018 Sep 17;18(4):139-141. doi: 10.1016/j.tjem.2018.09.001. eCollection 2018 Dec.

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验