Bronchoscopic protected catheter brush for the diagnosis of pulmonary infections.

作者信息

Marquette C H, Ramon P, Courcol R, Wallaert B, Tonnel A B, Voisin C

机构信息

Department of Pneumologie, Hopital A. Calmette, Lille, France.

出版信息

Chest. 1988 Apr;93(4):746-50. doi: 10.1378/chest.93.4.746.

Abstract

A new bronchoscopic-protected catheter brush (BPCB), designed to obtain uncontaminated bronchial secretions, was studied in vitro and in vivo. The device was composed of a standard biopsy brush, protected by a single catheter and occluded with an agar plug. Ejection of the plug was obtained neither by advancing the brush nor by advancing an inner cannula (as in a telescoping catheter brush), but instead, by an air flux, provided by a syringe which was connected to the proximal tip. In the first part of the study the ability of the BPCB to obtain uncontaminated specimens was tested in comparison with the reference telescoping catheter brush (BFW brush 10/70/90, Medi-Tech Corp Watertown, MA). Catheters of each type were successively passed through the inner channel of a bronchofiberscope which was contaminated with Klebsiella pneumonia. After ejection of the distal plug, sampling of bronchial secretions infected with a marker organism (Pseudomonas aeruginosa), was performed with the brush. Culture of brush specimens of each type of catheter grew the marker organism in pure culture and obtained the same amount of bronchial secretions (0.001 ml). The manual vortexing of the brush in the transport medium (Ringer's solution) proved to be as effective as the mechanical vortexing so that transecting of the brush was no longer mandatory. In the second part of this study, paired bronchial samplings from 27 patients were performed using both types of catheters and similar results for both were obtained. In these in vitro studies, completed by a clinical trial, our single-sheathed, plugged catheter brush proved to be as reliable as the double telescoping catheter brush. However, because of its relatively simple conception, making it easier to use and lower in cost than the double catheter brush, routine use of this sampling device should be considered.

摘要

一种旨在获取未受污染支气管分泌物的新型支气管镜保护导管刷(BPCB)在体外和体内进行了研究。该装置由一个标准活检刷组成,由单个导管保护并用琼脂塞封闭。琼脂塞的排出既不是通过推进刷子也不是通过推进内套管(如伸缩式导管刷那样),而是通过连接到近端尖端的注射器提供的气流来实现。在研究的第一部分,将BPCB获取未受污染标本的能力与参考伸缩式导管刷(BFW刷10/70/90,Medi-Tech公司,马萨诸塞州沃特敦)进行了比较测试。每种类型的导管依次通过被肺炎克雷伯菌污染的支气管纤维镜的内部通道。在排出远端塞子后,用刷子对感染标记生物(铜绿假单胞菌)的支气管分泌物进行采样。每种类型导管的刷子标本培养均在纯培养物中培养出标记生物,并且获得了相同量的支气管分泌物(0.001毫升)。事实证明,在运输培养基(林格氏溶液)中手动涡旋刷子与机械涡旋一样有效,因此不再需要切断刷子。在本研究的第二部分,对27例患者使用两种类型的导管进行了配对支气管采样,两者获得了相似的结果。在这些由临床试验完成的体外研究中,我们的单鞘、带塞导管刷被证明与双伸缩导管刷一样可靠。然而,由于其概念相对简单,使其比双导管刷更易于使用且成本更低,因此应考虑常规使用这种采样装置。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验