A Compact Multi-Distance DCS and Time Domain NIRS Hybrid System for Hemodynamic and Metabolic Measurements.

机构信息

Dipartimento di Fisica, Politecnico di Milano, piazza Leonardo da Vinci 32, 20133 Milan, Italy.

Dipartimento di Elettronica, Informazione e Bioingegneria, Politecnico di Milano, via Ponzio 34/5, 20133 Milan, Italy.

出版信息

Sensors (Basel). 2021 Jan 28;21(3):870. doi: 10.3390/s21030870.

Abstract

In this work, we present a new multi-distance diffuse correlation spectroscopy (DCS) device integrated with a compact state-of-the-art time domain near infrared spectroscopy (TD-NIRS) device. The hybrid DCS and TD-NIRS system allows to retrieve information on blood flow, tissue oxygenation, and oxygen metabolic rate. The DCS device performances were estimated in terms of stability, repeatability, ability in retrieving variations of diffusion coefficient, influence of the tissue optical properties, effect of varying count rates and depth sensitivity. Crosstalk between DCS and TD-NIRS optical signals was also evaluated. Finally, in vivo experiments (venous and arterial cuff occlusions on the arm) were conducted to test the ability of the hybrid system in measuring blood flow variations.

摘要

在这项工作中,我们展示了一种新的多距离漫射相关光谱(DCS)设备,该设备与紧凑型最先进的时域近红外光谱(TD-NIRS)设备集成在一起。混合 DCS 和 TD-NIRS 系统可用于获取血流、组织氧合和氧代谢率的信息。DCS 设备的性能是根据稳定性、重复性、获取扩散系数变化的能力、组织光学特性的影响、计数率和深度灵敏度变化的影响来评估的。还评估了 DCS 和 TD-NIRS 光学信号之间的串扰。最后,进行了体内实验(手臂上的静脉和动脉袖带阻塞),以测试混合系统测量血流变化的能力。