The goal of this research is to compare the intraocular pressure (IOP) and the mean central corneal thickness (CCT) values obtained from the measurements with Nidek NT-530P and Canon TX-20P devices with the values obtained by ultrasound pachymetry (UP) and Goldmann applanation tonometry (GAT) in children.
This prospective study was conducted with 119 healthy children. The measurements were repeated three times for each eye. The intraclass correlation coefficient (ICC) was used to assess the correlation between the measurements obtained from different devices. The Bland-Altman plot was used to analyze the agreement between two different devices graphically. The measurements were taken in the same order in all subjects: TX-20P, NT-530P, UP, and GAT.
The mean age of the children was 10.1 ± 3.2 (6-17) years. The mean CCT values for the eyes were 568.90 and 569.68 (TX-20P), 571.44 and 566.37 (NT-530P), 564.77 and 564.67 μm UP, (right and left, respectively). The highest correspondence observed for CCT was between UP and NT-530P devices (ICC, 0.982, the left eye). The mean IOP measurements for Canon TX-20P, NT-530P, and GAT were 16.5 ± 3.2, 16.3 ± 3.2, and 16.8 ± 3.6 mmHg for the right eyes, respectively, while those for the left eyes were 16.6 ± 3.1, 16.2 ± 3.3, and 16.8 ± 3.7 mmHg. The highest correspondence was noted between measurements obtained using GAT and Nidek NT-530P devices (ICC, 0.945, in the left eye).
Tono/pachymetry devices in the healthy children are provides significant agreements with the gold standard methods for IOP and CCT measurements.Clinical Trial Registration number: 2016-94.
本研究旨在比较 Nidek NT-530P 和 Canon TX-20P 设备与超声角膜测厚仪(UP)和 Goldmann 压平眼压计(GAT)测量值的眼压(IOP)和平均中央角膜厚度(CCT)值,并将其与儿童的测量值进行比较。
本前瞻性研究纳入了 119 名健康儿童。每只眼重复测量 3 次。采用组内相关系数(ICC)评估不同设备测量值之间的相关性。采用 Bland-Altman 图以图形方式分析两种不同设备之间的一致性。所有受试者均按 TX-20P、NT-530P、UP 和 GAT 的相同顺序进行测量。
儿童的平均年龄为 10.1±3.2(6-17)岁。右眼 TX-20P 和 NT-530P 的 CCT 值分别为 568.90 和 569.68μm,左眼分别为 571.44 和 566.37μm;右眼 UP 和左眼 UP 的 CCT 值分别为 564.77 和 564.67μm。CCT 值的相关性最高的是 UP 与 NT-530P 设备(ICC,0.982,左眼)。右眼 Canon TX-20P、NT-530P 和 GAT 的平均 IOP 测量值分别为 16.5±3.2、16.3±3.2 和 16.8±3.6mmHg,左眼分别为 16.6±3.1、16.2±3.3 和 16.8±3.7mmHg。GAT 与 Nidek NT-530P 设备测量值的相关性最高(ICC,0.945,左眼)。
在健康儿童中,眼压/角膜测厚仪与 IOP 和 CCT 测量的金标准方法具有显著的一致性。