Department of Rehabilitation, Orthoptics and Visual Science Course, School of Allied Health Sciences, Kitasato University, 1-15-1 Kitasato, Minamiku, Sagamihara, Kanagawa 252-0373, Japan ; Department of Ophthalmology, Kitasato University Hospital, 1-15-1 Kitasato, Minamiku, Sagamihara, Kanagawa 252-0373, Japan.

出版信息

Biomed Res Int. 2013;2013:370592. doi: 10.1155/2013/370592. Epub 2013 Oct 8.

Abstract

PURPOSE

To investigate the repeatability and reproducibility of intraocular pressure (IOP) and central corneal thickness (CCT) measurements using a noncontact tono/pachymeter (NT-530P) and to assess the correlation of CCT with IOP.

METHODS

Forty-six eyes of healthy volunteers were measured by two examiners. Three consecutive measurements per eye were performed. Repeatability was assessed using the coefficient of variation, and reproducibility was assessed using Bland-Altman plots. Linear correlations were used to determine agreement between CCT and noncorrected IOP and CCT and corrected IOP, which was calculated using a formula built into the NT-530P.

RESULTS

The coefficient of variation for IOP was 6.4% and for CCT was 0.4%. The 95% limits of agreement between examiners were -0.17 ± 1.42 mmHg (range: -2.95 to 2.61 mmHg) for IOP, -0.93 ± 4.37  μ m (range: -9.50 to 7.64  μm) for CCT. The corrected IOP was significantly higher than the noncorrected IOP (P = 0.010.3). The noncorrected IOP significantly correlated with CCT (r = -0.4883, P = 0.0006). The corrected IOP showed no significant correlation with CCT (r = -0.0285, P = 0.8509).

CONCLUSIONS

NT-530P offered repeatability and reproducibility in both IOP and CCT measurements. The corrected IOP calculated using the NT-530P was independent of the CCT, suggesting that this IOP may be less influenced by the central corneal thickness.

摘要

目的

使用非接触眼压计(NT-530P)研究眼压(IOP)和中央角膜厚度(CCT)测量的重复性和再现性,并评估 CCT 与 IOP 的相关性。

方法

对 46 例健康志愿者的 46 只眼进行了测量,由两名检查者进行。每只眼进行 3 次连续测量。使用变异系数评估重复性,使用 Bland-Altman 图评估再现性。使用线性相关性来确定 CCT 与未经校正的 IOP 之间以及 CCT 与通过 NT-530P 内置公式计算得出的校正 IOP 之间的一致性。

结果

IOP 的变异系数为 6.4%,CCT 的变异系数为 0.4%。IOP 检查者之间的 95%一致性界限为-0.17±1.42mmHg(范围:-2.95 至 2.61mmHg),CCT 的为-0.93±4.37μm(范围:-9.50 至 7.64μm)。校正后的 IOP 明显高于未经校正的 IOP(P=0.010.3)。未经校正的 IOP 与 CCT 显著相关(r=-0.4883,P=0.0006)。校正后的 IOP 与 CCT 无显著相关性(r=-0.0285,P=0.8509)。

结论

NT-530P 在 IOP 和 CCT 测量中均具有重复性和再现性。使用 NT-530P 计算得出的校正 IOP 与 CCT 无关,这表明该 IOP 可能较少受到中央角膜厚度的影响。