Axial length acquisition success rates and agreement of two swept-source optical biometers in eyes with dense cataracts.

作者信息

Orts-Vila Paz, Tañá-Sanz Santiago, Tello-Elordi Cristina, Montés-Micó Robert, Tañá-Rivero Pedro

机构信息

Oftalvist, Alicante, Spain.

Department of Optics and Optometry and Vision Sciences, University of Valencia, Valencia, Spain.

出版信息

Front Med (Lausanne). 2024 Sep 25;11:1449867. doi: 10.3389/fmed.2024.1449867. eCollection 2024.

Abstract

BACKGROUND

Swept-source optical coherence tomography-based (SS-OCT) biometers have been used in different clinical studies with the aim of assessing the accuracy of the technique, specifically in eyes with dense cataracts. Our objective is to evaluate the axial length acquisition success rates and agreement of two SS-OCT biometers when measuring axial length and biometric parameters in eyes with dense cataracts.

METHODS

46 eyes (46 patients) with dense cataracts (LOCS III grade ≥ 4) were measured 3 consecutive times using the Eyestar 900 and Argos SS-OCT biometers. Keratometry (K1, flat and K2, steep), central corneal thickness (CCT), white-to-white (WTW), anterior-chamber-depth (ACD), lens-thickness (LT), and axial length were measured using both biometers. The percentage acquisition success rate and a Bland-Altman analysis to determine the agreement between the biometers were calculated. Corrected and uncorrected distance visual acuity, subjective refraction, and axial length (to assess preoperative axial length accuracy) were measured 1-month post-cataract surgery.

RESULTS

The mean LOCS III score was 4.37 ± 0.68. The acquisition success rates for both biometers was 100%. There were statistically significant differences between the two SS-OCT biometers for all parameters evaluated ( < 0.05). The mean differences for K1, K2, CCT, WTW, ACD, LT and axial length were 0.106 D, 0.128 D, -6.347 μm, -0.054 mm, 0.095 mm, 0.110 mm, and -0.036 mm, respectively. The mean pre- and post-surgery axial length difference was -0.036 mm for the Eyestar 900 and -0.020 mm for the Argos. This difference was ≤ 0.1 mm in 97.82% of eyes with the Eyestar 900 and in 100% of eyes with the Argos.

CONCLUSION

SS-OCT biometry successfully measures axial length in dense cataracts. The differences between biometers in some parameters may have a clinically significant impact but should be judged individually. The pre- and post-surgery axial length differences for each biometer can be considered clinically negligible and should not affect the IOL power calculation.

摘要

背景

基于扫频源光学相干断层扫描(SS-OCT)的生物测量仪已用于不同的临床研究,旨在评估该技术的准确性,尤其是在患有致密性白内障的眼睛中。我们的目的是评估两种SS-OCT生物测量仪在测量致密性白内障眼睛的眼轴长度和生物测量参数时的眼轴长度采集成功率及一致性。

方法

使用Eyestar 900和Argos SS-OCT生物测量仪对46例(46只眼)患有致密性白内障(LOCS III级≥4)的患者进行连续3次测量。使用两种生物测量仪测量角膜曲率(K1,平曲率和K2,陡曲率)、中央角膜厚度(CCT)、白对白(WTW)、前房深度(ACD)、晶状体厚度(LT)和眼轴长度。计算采集成功率百分比,并进行Bland-Altman分析以确定两种生物测量仪之间的一致性。在白内障手术后1个月测量矫正和未矫正的远视力、主观验光和眼轴长度(以评估术前眼轴长度准确性)。

结果

平均LOCS III评分为4.37±0.68。两种生物测量仪的采集成功率均为100%。在所有评估参数方面,两种SS-OCT生物测量仪之间存在统计学显著差异(<0.05)。K1、K2、CCT、WTW、ACD、LT和眼轴长度的平均差异分别为0.106 D、0.128 D、-6.347μm、-0.054 mm、0.095 mm、0.110 mm和-0.036 mm。Eyestar 900的术前和术后眼轴长度平均差异为-0.036 mm,Argos为-0.020 mm。在使用Eyestar 900的眼睛中,97.82%的差异≤0.1 mm,在使用Argos的眼睛中,100%的差异≤0.1 mm。

结论

SS-OCT生物测量法可成功测量致密性白内障的眼轴长度。生物测量仪在某些参数上的差异可能具有临床显著影响,但应个别判断。每种生物测量仪的术前和术后眼轴长度差异在临床上可忽略不计,不应影响人工晶状体屈光度计算。