Accuracy of Digital Impressions Obtained Using Six Intraoral Scanners in Partially Edentulous Dentitions and the Effect of Scanning Sequence.

出版信息

Int J Prosthodont. 2021 Jan-Feb;34(1):101-108. doi: 10.11607/ijp.6834.

Abstract

PURPOSE

To compare the accuracy of six intraoral scanners in two different partially edentulous maxillary models and to evaluate the effect of scanning sequence on accuracy.

MATERIALS AND METHODS

Maxillary Kennedy Class I and Class IV situations were used as reference models. The reference datasets were obtained by scanning the models using a highly accurate industrial scanner (ATOS Core 80, GOM). The following six intraoral scanners were evaluated: Trios 3 (3Shape), iTero Element 2 (Align Technology), Emerald (Planmeca), CEREC Omnicam (Dentsply Sirona), CEREC Primescan (Dentsply Sirona), and Virtuo Vivo (Dental Wings). A total of 120 scans from both models were obtained using the six intraoral scanners and divided into two groups based on scanning sequence. Accuracy was evaluated by deviation analysis using 3D image processing software (Geomagic Studio 12, 3D Systems). Kruskal Wallis and Mann-Whitney U tests were performed (P ≤ .05) for statistical analysis.

RESULTS

There were significant differences in the accuracy of digital impressions among intraoral scanners and scanning sequences. The trueness of the Trios scanner and the precision of the Trios, Primescan, and iTero scanners were significantly higher than for the other scanners. The Emerald had the lowest accuracy among the six intraoral scanners tested. Accuracy was affected by scanning sequence when using the Virtuo Vivo, Emerald, Primescan, and iTero.

CONCLUSION

In Kennedy Class I and Class IV partially edentulous cases, it is useful to consider that the intraoral scanner used may affect the accuracy of the digital impression.

摘要

目的

比较 6 种口内扫描仪在 2 种不同的部分无牙颌上颌模型中的准确性,并评估扫描顺序对准确性的影响。

材料与方法

采用 Kennedy 类Ⅰ和类Ⅳ上颌模型作为参考模型。参考数据集通过使用高精度工业扫描仪(ATOS Core 80,GOM)扫描模型获得。评估了以下 6 种口内扫描仪:Trios 3(3Shape)、iTero Element 2(Align Technology)、Emerald(Planmeca)、CEREC Omnicam(Dentsply Sirona)、CEREC Primescan(Dentsply Sirona)和 Virtuo Vivo(Dental Wings)。使用这 6 种口内扫描仪从两个模型中总共获得 120 个扫描,并根据扫描顺序将其分为两组。使用 3D 图像处理软件(Geomagic Studio 12,3D Systems)通过偏差分析评估准确性。进行 Kruskal Wallis 和 Mann-Whitney U 检验(P≤.05)进行统计分析。

结果

口内扫描仪和扫描顺序对数字印模的准确性有显著影响。Trios 扫描仪的准确性和 Trios、Primescan 和 iTero 扫描仪的精度明显高于其他扫描仪。在测试的 6 种口内扫描仪中,Emerald 的准确性最低。当使用 Virtuo Vivo、Emerald、Primescan 和 iTero 时,扫描顺序会影响准确性。

结论

在 Kennedy 类Ⅰ和类Ⅳ部分无牙颌病例中,需要考虑使用的口内扫描仪可能会影响数字印模的准确性。