Evaluation of the DINAMAP blood pressure monitor in an ambulatory primary care setting.

作者信息

Ornstein S, Markert G, Litchfield L, Zemp L

机构信息

Department of Family Medicine, Medical University of South Carolina, Charleston 29425-5820.

出版信息

J Fam Pract. 1988 May;26(5):517-21.

PMID:3367116
Abstract

Automatic blood pressure recorders have gained acceptance in many clinical settings. New devices have usually been validated with invasive monitoring as the "gold standard." There is a lack of sound empirical evidence, however, supporting the routine use of these monitors in ambulatory settings. This study evaluated the DINAMAP 8100, an oscillometric automated blood pressure monitor, using the Hawksley Random-Zero Sphygmomanometer as the standard. A sample of 80 normotensive and hypertensive ambulatory patients from the Department of Family Medicine at the Medical University of South Carolina were studied. A clinical trial was conducted in which readings from the DINAMAP 8100 were compared with those from the Hawksley Random-Zero Sphygmomanometer, in a 2 (instrument) X 2 (arm) X 2 (investigators) X 4 (pairs of simultaneous measurements) factorial design. The DINAMAP 8100 overestimated systolic readings (mean difference = 7.6 +/- 9.1 mmHg, P less than .0001, paired t test). More than one third of systolic measurements and one quarter of diastolic measurements were greater than 10 mmHg discrepant from the standard. The results of this study suggest that routine use of the DINAMAP 8100 would lead to serious misclassification errors in screening for hypertension and in the follow-up of known hypertensive patients. The DINAMAP 8100, therefore, is not an appropriate instrument for routine use in primary care settings.

摘要

自动血压记录仪已在许多临床环境中得到认可。新设备通常已通过侵入性监测作为“金标准”进行了验证。然而,缺乏可靠的经验证据支持在门诊环境中常规使用这些监测仪。本研究以Hawksley随机零位血压计作为标准,对DINAMAP 8100型示波法自动血压监测仪进行了评估。研究对象为南卡罗来纳医科大学家庭医学系的80名血压正常和高血压的门诊患者。进行了一项临床试验,采用2(仪器)×2(手臂)×2(研究者)×4(同步测量对)析因设计,将DINAMAP 8100的读数与Hawksley随机零位血压计的读数进行比较。DINAMAP 8100高估了收缩压读数(平均差异=7.6±9.1 mmHg,P<0.0001,配对t检验)。超过三分之一的收缩压测量值和四分之一的舒张压测量值与标准值相差超过10 mmHg。本研究结果表明,常规使用DINAMAP 8100会在高血压筛查和已知高血压患者的随访中导致严重的错误分类误差。因此,DINAMAP 8100不是基层医疗环境中常规使用的合适仪器。

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验