Suppr超能文献

mfERG 测试中的眼底运动。

Fundus motion during mfERG testing.

机构信息

University of Houston College of Optometry, Houston, TX, USA.

出版信息

Doc Ophthalmol. 2021 Oct;143(2):129-139. doi: 10.1007/s10633-021-09829-9. Epub 2021 Mar 13.

Abstract

PURPOSE

The purpose of this study was to assess eye movements during a multifocal ERG (mfERG) recording. This study evaluated the relationship between bivariate contour ellipse areas (BCEAs), mfERG amplitudes (Amps) and mfERG implicit times (ITs) with repeat testing and experienced subjects.

METHODS

Thirty subjects were selected (15 experienced to ocular procedures and 15 novices). All were confirmed to have healthy retinas and at least 20/25 vision. MfERGs with a stimulus near 100% contrast and 4-min m-sequence were recorded on two different days using our common clinical technique, which did not constrain the head. VERIS with fundus monitoring system was used for recording with a Burian-Allen electrode. An external camera captured the fundus during each mfERG recording. The optic nerve head position was tracked in each video using a custom algorithm in order to determine BCEAs. Each subject performed one mfERG on two different days. MfERGs were analyzed for Amps and ITs for the fovea and whole eye.

RESULTS

There was no correlation between the mfERG metrics and BCEAs with repeat testing. There were also no differences between the experienced and novice subjects for mfERG Amps, ITs or BCEAs. Eye movements between visits were highly correlated (multiple r = 0.67). BCEAs were larger during mfERGs (1.04 ± 0.8 deg) than those observed in previous literature using brief viewing tasks (< 0.3 deg). The proportion of time spent fixating within 1.0 and 2.0 degrees of the central hexagon was 68 and 93%, respectively.

CONCLUSIONS

This study is the first to evaluate the stability of the retina while recording a mfERG in healthy subjects and indicates that the center of fixation during a mfERG stays within the central hexagon. Eye stability during an initial recording is the best indicator of stability on the second recording. The amount of movement during these recordings did not seem to affect the mfERG Amps or ITs. These data suggest clinical confidence with mfERGs when recording novice patients.

摘要

目的

本研究旨在评估多焦视网膜电图(mfERG)记录过程中的眼动。本研究评估了双变量轮廓椭圆面积(BCEA)、mfERG 幅度(Amps)和 mfERG 潜伏期(ITs)与重复测试和有经验的受试者之间的关系。

方法

选择 30 名受试者(15 名对眼部手术有经验,15 名无经验)。所有受试者均被证实视网膜健康,视力至少为 20/25。使用我们常见的临床技术,在两天内记录刺激接近 100%对比度和 4 分钟 m 序列的 mfERG,该技术不限制头部。使用带有眼底监测系统的 VERIS 记录,使用 Burian-Allen 电极进行记录。在每次 mfERG 记录过程中,外部相机捕获眼底图像。使用自定义算法在每个视频中跟踪视盘的位置,以确定 BCEA。每个受试者在两天内进行一次 mfERG。分析 fovea 和整个眼睛的 mfERG 幅度和 ITs。

结果

重复测试时,mfERG 指标与 BCEA 之间没有相关性。在 mfERG Amps、ITs 或 BCEA 方面,有经验和无经验的受试者之间也没有差异。两次就诊之间的眼动高度相关(多重 r=0.67)。mfERG 期间的 BCEA 比以前使用短暂观察任务观察到的文献中的 BCEA 更大(<0.3 度)(1.04±0.8 度)。在中央六边形内 1.0 和 2.0 度范围内注视的时间比例分别为 68%和 93%。

结论

本研究首次评估了健康受试者记录 mfERG 时视网膜的稳定性,并表明 mfERG 期间的注视中心保持在中央六边形内。初次记录时的眼稳定性是第二次记录时稳定性的最佳指标。这些记录过程中的运动幅度似乎不会影响 mfERG Amps 或 ITs。这些数据表明,在记录新手患者时,对 mfERG 有临床信心。