Stuberg W A, Metcalf W K
C. Louis Meyer Children's Rehabilitation Institute, University of Nebraska Medical Center, Omaha 68131.
Phys Ther. 1988 Jun;68(6):977-82. doi: 10.1093/ptj/68.6.977.
The purpose of this study was to examine intratester and test-retest reliability using a hand-held dynamometer for the measurement of isometric muscle strength in 28 healthy children and children with Duchenne muscular dystrophy. The Dystrophic Group consisted of 14 children diagnosed with Duchenne muscular dystrophy, and the Healthy Group consisted of 14 age-matched children with no history of orthopedic or neuromuscular disorders. One physical therapist tested hip and knee extension, elbow flexion, and shoulder abduction in each child bilaterally. A two-way analysis of variance for repeated measures was used to analyze differences between measurements taken within and across the testing sessions. Pearson product-moment correlation coefficients were determined on mean values across the testing sessions for each variable. No significant differences (p greater than .05) between measurements taken within or across testing sessions were found in either the Dystrophic Group or the Healthy Group. Correlation coefficients for the Dystrophic Group ranged from .83 to .99 for the variables tested. Correlation coefficients for the Healthy Group ranged from .74 to .99. The results suggest that the hand-held dynamometer can be used as a reliable instrument in measuring the isometric strength of selected muscles in children.
本研究的目的是使用手持测力计,检测28名健康儿童和杜氏肌营养不良症患儿等长肌肉力量的测试者内和重测信度。营养不良组由14名被诊断为杜氏肌营养不良症的儿童组成,健康组由14名年龄匹配且无骨科或神经肌肉疾病史的儿童组成。一名物理治疗师对每个儿童双侧的髋关节和膝关节伸展、肘关节屈曲和肩关节外展进行了测试。采用重复测量的双向方差分析来分析测试期间内和测试期间之间测量值的差异。对每个变量在测试期间的平均值确定Pearson积差相关系数。在营养不良组或健康组中,测试期间内或测试期间之间的测量值均未发现显著差异(p大于0.05)。营养不良组测试变量的相关系数范围为0.83至0.99。健康组的相关系数范围为0.74至0.99。结果表明,手持测力计可作为测量儿童特定肌肉等长力量的可靠工具。