Long-term inter-platform reproducibility, bias, and linearity of commercial PDFF MRI methods for fat quantification: a multi-center, multi-vendor phantom study.

机构信息

Imaging Institute, Cleveland Clinic, 9500 Euclid Avenue, A21, Cleveland, OH, 44195, USA.

Glickman Urological and Kidney Institute, Cleveland Clinic, 9500 Euclid Avenue, Cleveland, OH, 44195, USA.

出版信息

Eur Radiol. 2021 Oct;31(10):7566-7574. doi: 10.1007/s00330-021-07851-8. Epub 2021 Mar 25.

Abstract

OBJECTIVES

Proton density fat fraction (PDFF) is a validated biomarker of tissue fat quantification. However, validation has been limited to single-center or multi-center series using non-FDA-approved software. Thus, we assess the bias, linearity, and long-term reproducibility of PDFF obtained using commercial PDFF packages from several vendors.

METHODS

Over 35 months, 438 subjects and 16 volunteers from a multi-center observational trial underwent PDFF MRI measurements using a 3-T MR system from one of three different vendors or a 1.5-T system from one vendor. Fat-water phantom sets were measured as part of each subject's examination. Manual region-of-interest measurements on the %fat image, then cross-sectional bias, linearity, and long-term reproducibility were assessed.

RESULTS

Three hundred ninety-two phantom measurements were evaluable (90%). Bias ranged from 2.4 to - 3.8% for the lowest to the highest weight %fat. Regression fits of PDFF against synthesis weight %fat showed negligible non-linear effects and a linear slope of 0.94 (95% confidence interval: 0.938, 0.947). We observed significant vendor (p < 0.001) and field strength (p < 0.001) differences in bias and longitudinal variability. When the results were pooled across sites, vendors, and field strengths, the estimated reproducibility coefficient was 6.93% (95% CI: 6.25%, 7.81%).

CONCLUSIONS

This study demonstrated good linearity, accuracy, and reproducibility for all investigated manufacturers and field strengths. However, significant vendor-dependent and field strength-dependent bias were found. While longitudinal PDFF measurements may be made using different field strength or vendor MR systems, if the MR system is not the same, based on these results, only PDFF changes ≥ 7% can be considered a true difference.

KEY POINTS

• Phantom fat fraction (PDFF) MRI measurements over 35 months demonstrated good linearity, accuracy, and reproducibility for the vendor systems investigated. • Non-linear effects were negligible (linear slope of 0.94) over 0-100% fat; however, significant vendor (p < 0.001) and field strength (p<0.001) differences in bias and longitudinal variability were identified. Bias ranged from 2.4 to - 3.8% for 0-100 weight% fat, respectively. • Measurement bias could affect the accuracy of PDFF in clinical use. As the reproducibility coefficient was 6.93%, only greater changes in % fat can be considered true differences when making longitudinal PDFF measurements on different MR systems.

摘要

目的

质子密度脂肪分数(PDFF)是组织脂肪定量的一种经过验证的生物标志物。然而,这种验证仅限于使用非 FDA 批准的软件的单中心或多中心系列。因此,我们评估了来自多个供应商的商业 PDFF 软件包获得的 PDFF 的偏倚、线性和长期可重复性。

方法

在 35 个月的时间里,来自一项多中心观察性试验的 438 名受试者和 16 名志愿者在一台来自三个不同供应商的 3T MR 系统或一台来自一个供应商的 1.5T 系统上进行了 PDFF MRI 测量。脂肪水体模套件作为每个受试者检查的一部分进行测量。在%fat 图像上手动进行感兴趣区域测量,然后评估截面偏差、线性度和长期可重复性。

结果

392 个体模测量结果可评估(90%)。对于最低和最高体重%脂肪,偏差范围为 2.4%至-3.8%。PDFF 与合成体重%脂肪的回归拟合显示出可忽略的非线性影响和 0.94 的线性斜率(95%置信区间:0.938,0.947)。我们观察到供应商(p<0.001)和场强(p<0.001)对偏差和纵向变异性的显著差异。当结果在站点、供应商和场强之间进行汇总时,估计的重现性系数为 6.93%(95%CI:6.25%,7.81%)。

结论

本研究证明了所有研究的制造商和场强的良好线性、准确性和可重复性。然而,发现了显著的供应商依赖性和场强依赖性偏差。虽然可以使用不同场强或供应商的 MR 系统进行纵向 PDFF 测量,但根据这些结果,如果 MR 系统不同,只有≥7%的 PDFF 变化才能被认为是真正的差异。

要点

• 在 35 个月的时间里,对幻影脂肪分数(PDFF)MRI 测量进行了研究,证明了所研究的供应商系统具有良好的线性、准确性和可重复性。• 在 0-100%脂肪范围内,非线性影响可忽略不计(线性斜率为 0.94);然而,发现了显著的供应商(p<0.001)和场强(p<0.001)对偏差和纵向变异性的差异。偏差范围分别为 0-100 重量%脂肪的 2.4%至-3.8%。• 测量偏差可能会影响 PDFF 在临床使用中的准确性。由于重现性系数为 6.93%,只有在不同的 MR 系统上进行纵向 PDFF 测量时,%脂肪的更大变化才能被认为是真正的差异。