To measure the 24-h intraocular pressure (IOP) by Icare PRO rebound in healthy and primary open-angle glaucoma (POAG) eyes and compare it with non-contact tonometry (NCT).
Thirty POAG patients, who were under IOP-lowering treatment, and 30 healthy subjects were included. Participants were hospitalized overnight for the 24-h IOP measurement. IOPs were measured by Icare PRO and NCT according to a standard protocol every 2 h during 24 h. The 24-h IOP curve and IOP-related parameters were compared between Icare PRO and NCT groups in POAG and healthy eyes.
The IOPs measured by Icare PRO in habitual position increased notably at 22:00 in the normal group and at 20:00 in the POAG group, reached peak at 0:00, stayed high until 4:00, and then decreased in both groups (all p < 0.05). The POAG patients had higher mean 24-h IOP, peak IOP, IOP fluctuation, and greater IOP change from supine to sitting position in the nocturnal period than those in the normal subjects even after adjusting for eyes, age, gender, CCT, and axial length (all p < 0.05).
The Icare PRO provides a well-tolerated approach for 24-h IOP monitoring in habitual position. Twenty-four-hour IOP in habitual position is more sensitive for detecting high nocturnal IOP peaks and greater IOP fluctuation for POAG patients.
通过 Icare PRO 回弹测量健康人和原发性开角型青光眼(POAG)眼的 24 小时眼压(IOP),并将其与非接触眼压计(NCT)进行比较。
纳入 30 名正在接受降眼压治疗的 POAG 患者和 30 名健康受试者。参与者被安排住院过夜以进行 24 小时眼压测量。按照标准方案,每 2 小时通过 Icare PRO 和 NCT 测量 IOP,在 24 小时内进行测量。比较了 POAG 眼和健康眼中 Icare PRO 和 NCT 组的 24 小时眼压曲线和眼压相关参数。
正常组习惯性体位的 IOP 在 22:00 明显升高,POAG 组在 20:00 升高,在 0:00 达到峰值,一直持续到 4:00,然后两组均下降(均 P<0.05)。POAG 患者夜间平均 24 小时 IOP、峰值 IOP、IOP 波动以及从仰卧位到坐姿时的眼压变化均高于正常组,即使在调整了眼别、年龄、性别、CCT 和眼轴长度后(均 P<0.05)。
Icare PRO 为习惯性体位 24 小时眼压监测提供了一种耐受性良好的方法。习惯性体位 24 小时眼压更能敏感地检测到 POAG 患者夜间高眼压峰值和更大的眼压波动。