Suppr超能文献

7T超高分辨率T2加权图像中运动及回顾性运动校正对血管周围间隙可视化和定量分析的影响

Effects of motion and retrospective motion correction on the visualization and quantification of perivascular spaces in ultrahigh resolution T2-weighted images at 7T.

作者信息

Zong Xiaopeng, Nanavati Soham, Hung Sheng-Che, Li Tengfei, Lin Weili

机构信息

Department of Radiology, University of North Carolina, Chapel Hill, North Carolina, USA.

Biomedical Research Imaging Center, University of North Carolina, Chapel Hill, North Carolina, USA.

出版信息

Magn Reson Med. 2021 Oct;86(4):1944-1955. doi: 10.1002/mrm.28847. Epub 2021 May 19.

Abstract

PURPOSE

Motion can strongly affect MRI image quality and derived imaging measures. We studied the effects of motion and retrospective motion correction (MC) on the visualization and quantitative measures of the perivascular space and penetrating vessel (PVSV) complex, an essential part of the glymphatic system, on high-resolution T -weighted MRI images at 7T.

METHODS

MC was achieved by adjusting k-space data based on head positions measured using fat navigator images. PVSV visibility and quantitative measures including diameter, volume fraction (VF), count, and contrast were compared between images with and without MC.

RESULTS

Without MC, VF, and count decreased significantly with increasing head rotation. MC improved PVSV visualization in all cases with severe motion artifacts. MC decreased diameter in white matter (WM) and increased VF, count, and contrast in basal ganglia and WM. The changes of VF, count, and contrast after MC strongly correlated with motion severity. MC eliminated the significant dependences of VF and count on rotation and reduced the inter-subject variations of VF and count. The effect sizes of age and breathing gas effects on VF and count, and contrast increased in most cases after MC, while those on diameter exhibited inconsistent behavior.

CONCLUSIONS

Motion affects PVSV quantification without MC. MC improves PVSV visibility and increases the statistical power of detecting physiological PVSV VF, count, and contrast changes but may have limited benefits for increasing the power for detecting diameter changes.

摘要

目的

运动可强烈影响MRI图像质量及衍生的成像测量值。我们研究了运动及回顾性运动校正(MC)对血管周围间隙和穿通血管(PVSV)复合体(脑淋巴系统的重要组成部分)在7T高分辨率T加权MRI图像上的可视化及定量测量的影响。

方法

通过基于使用脂肪导航图像测量的头部位置调整k空间数据来实现MC。比较有和没有MC的图像之间的PVSV可见性及包括直径、体积分数(VF)、数量和对比度在内的定量测量值。

结果

没有MC时,VF和数量随着头部旋转增加而显著降低。在所有存在严重运动伪影的情况下,MC均改善了PVSV的可视化。MC使白质(WM)中的直径减小,并使基底神经节和WM中的VF、数量及对比度增加。MC后VF、数量和对比度的变化与运动严重程度密切相关。MC消除了VF和数量对旋转的显著依赖性,并减少了VF和数量的个体间差异。在大多数情况下,MC后年龄和呼吸气体对VF、数量及对比度的效应大小增加,而对直径的效应大小表现出不一致的行为。

结论

没有MC时,运动影响PVSV定量。MC改善了PVSV的可见性,并增加了检测生理性PVSV的VF、数量和对比度变化的统计功效,但对增加检测直径变化的功效可能益处有限。