Kommineni Usha Bhanu, Mohamed Ashik, Vaddavalli Pravin K, Reddy Jagadesh C
Cataract and Refractive Surgery Services, Cornea Institute, L V Prasad Eye Institute, Hyderabad, India.
Department of Ophthalmic Biophysics, L V Prasad Eye Institute, Hyderabad, India.
Eur J Ophthalmol. 2022 May;32(3):1496-1503. doi: 10.1177/11206721211020633. Epub 2021 May 29.
To compare the total keratometry (TK) and astigmatism measurements in eyes with cataract using automated keratometry of sweptsource optical coherence tomography (ss-OCT), optical low-coherence reflectometry (OLCR), simulated keratometry (SimK), and total corneal power (TCP) of combined placidodual Scheimpflug imaging system.
The study was conducted at LV Prasad Eye Institute, Hyderabad, India.
Retrospective evaluation of electronic medical records of patients who were evaluated for cataract surgery.
Twenty-eight eyes of 28 patients were included in the study. All patients evaluated for cataract surgery underwent corneal power measurements using three devices: ssOCT, OLCR, and combined placidodual Scheimpflug imaging were included in the study. Vector analysis was performed to evaluate corneal astigmatism and Bland-Altman analysis was conducted to evaluate the limits of agreement of similar parameters among devices.
The mean TK was statistically significantly different from the keratometry obtained from optical biometers and values measured by the Scheimpflug imaging system. The magnitude of mean difference was greater between TK and TCP (0.75 ± 0.25) compared to other variables. The mean difference in astigmatism between TK, ss-OCT-K (0.09 ± 0.12, = 0.48), OCLR-K (0.10 ± 0.48, = 0.91), and TCP (0.09 ± 0.47, = 0.31) was not statistically significant but was statistically significant between TK and SimK values (0.23D ± 0.49D). The axis of orientation (<20°) of astigmatism was comparable (100%, 28 eyes) between two keratometry variables measured by ss-OCT.
There appears to be a greater correlation of automated keratometry, and TK values obtained from ss-OCT compared to other variables studied. The measurements from TK, Simk, and TCP cannot be used interchangeably.
使用扫频源光学相干断层扫描(ss-OCT)的自动角膜曲率测量、光学低相干反射测量(OLCR)、模拟角膜曲率测量(SimK)以及联合的普拉西多双像差Scheimpflug成像系统的总角膜屈光力(TCP),比较白内障患者眼睛的总角膜曲率(TK)和散光测量结果。
该研究在印度海得拉巴的LV普拉萨德眼科研究所进行。
对接受白内障手术评估的患者电子病历进行回顾性评估。
28例患者的28只眼睛纳入研究。所有接受白内障手术评估的患者均使用三种设备进行角膜屈光力测量:ssOCT、OLCR和联合的普拉西多双像差Scheimpflug成像均纳入研究。进行矢量分析以评估角膜散光,并进行Bland-Altman分析以评估各设备间相似参数的一致性界限。
平均TK与光学生物测量仪获得的角膜曲率测量值以及Scheimpflug成像系统测量的值在统计学上有显著差异。与其他变量相比,TK与TCP之间的平均差异幅度更大(0.75±0.25)。TK、ss-OCT-K(0.09±0.12,P = 0.48)、OCLR-K(0.10±0.48,P = 0.91)和TCP(0.09±0.47,P = 0.31)之间散光的平均差异无统计学意义,但TK与SimK值之间(0.23D±0.49D)有统计学意义。ss-OCT测量的两个角膜曲率变量之间散光的轴向(<20°)具有可比性(100%,28只眼)。
与所研究的其他变量相比,自动角膜曲率测量以及从ss-OCT获得的TK值之间似乎具有更强的相关性。TK、Simk和TCP的测量结果不能互换使用。