Killian Jacquelin M, Radin Rachel M, Gardner Cubby L, Kasuske Lalon, Bashirelahi Kylee, Nathan Dominic, Keyser David O, Cellucci Christopher J, Darmon David, Rapp Paul E
711th Human Performance Wing, Air Force Research Laboratory, Wright Patterson Air Force Base, WPAFB, OH 45433 USA.
Department of Psychiatry and Behavioral Medicine, Uniuversity of California, San Francisco, San Francisco, CA 94143, USA.
Behav Sci (Basel). 2021 May 2;11(5):68. doi: 10.3390/bs11050068.
Using healthy adult participants, seven measures of heart rate variability were obtained simultaneously from four devices in five behavioral conditions. Two devices were ECG-based and two utilized photoplethysmography. The 140 numerical values (measure, condition, device) are presented. The comparative operational reliability of the four devices was assessed, and it was found that the two ECG-base devices were more reliable than the photoplethysmographic devices. The interchangeability of devices was assessed by determining the between-device Limits of Agreement. Intraclass correlation coefficients were determined and used to calculate the standard error of measurement and the Minimal Detectable Difference. The Minimal Detectable Difference, MDD, quantifies the smallest statistically significant change in a measure and is therefore critical when HRV measures are used longitudinally to assess treatment response or disease progression.
使用健康成年参与者,在五种行为条件下从四种设备同时获取了七种心率变异性测量值。两种设备基于心电图,两种使用光电容积脉搏波描记法。呈现了140个数值(测量值、条件、设备)。评估了四种设备的相对操作可靠性,发现两种基于心电图的设备比光电容积脉搏波描记法设备更可靠。通过确定设备间的一致性界限来评估设备的互换性。确定了组内相关系数并用于计算测量标准误差和最小可检测差异。最小可检测差异(MDD)量化了测量中最小的统计学显著变化,因此当纵向使用心率变异性测量来评估治疗反应或疾病进展时至关重要。