Damia G, Cigada M, Solca M, Pelizzola A
Istituto di Anestesiologia e Rianimazione, Università degli Studi di Milano, Italy.
Intensive Care Med. 1988;14(1):60-3. doi: 10.1007/BF00254124.
A new model lung with the capacity for simulated spontaneous breathing is described. It consists of a modified commercial mechanical ventilator (Kontron ABT 4100), connected in parallel to a compliant system, a cylindric acrylic box with a latex thin membrane substituting for the top. Volume and compliance of the model are 2500 ml and 50 ml cmH2O-1, respectively. The modified ventilator simulates physiologic inspiratory flow at a rate of 10 to 30 min-1 and tidal volume up to 1000 ml, with an inspiratory to expiratory time ratio continuously variable between 1:4 and 4:1. The model has been tested under different respiratory assist techniques, connected either to continuous positive airway pressure proved to be reliable, versatile and bearing satisfactory resemblance to human ventilatory physiology.