Oculab眼压笔与Goldmann压平眼压计的临床比较。

Clinical comparison of the Oculab Tono-Pen to the Goldmann applanation tonometer.

作者信息

Kao S F, Lichter P R, Bergstrom T J, Rowe S, Musch D C

机构信息

Department of Ophthalmology, University of Michigan, Ann Arbor 48105.

出版信息

Ophthalmology. 1987 Dec;94(12):1541-4. doi: 10.1016/s0161-6420(87)33249-x.

Abstract

The Oculab Tono-Pen tonometer was compared with the Goldmann applanation tonometer through studies on 197 eyes of 103 patients with glaucoma or ocular hypertension to evaluate the reliability of the Tono-Pen in determining intraocular pressure (IOP). Patients were randomized into four groups and all applanations were done by the same investigator. The Tono-Pen initially used (Tono-Pen #1) failed to calibrate after being used on 77 eyes. Tono-Pen #2 was then obtained and tested. Each Tono-Pen showed high correlation with the Goldmann readings (r greater than or equal to 0.86 in both eyes). The overall sensitivity of the Tono-Pen in detecting IOPs of 21 mmHg or higher (as measured by the Goldmann instrument) was 62.1% in right eyes, 72.4% in left eyes, with corresponding specificities of 92.6% in the right and 97.1% in the left. Given the substantial increase in sensitivity noted with Tono-Pen #2, the Tono-Pen appears adequate for screening programs where an IOP of 21 mmHg or above is considered abnormal; however, at higher IOPs (greater than or equal to 30 mmHg) the Tono-Pens tended to underestimate Goldmann IOPs. At low IOPs (less than or equal to 9 mmHg) the Tono-Pens tended to overestimate the IOPs. Perhaps with further modifications of the instrument, it could be used for clinical diagnosis and management of glaucoma patients.

摘要

通过对103例青光眼或高眼压症患者的197只眼睛进行研究,将Oculab眼压笔眼压计与Goldmann压平眼压计进行比较,以评估眼压笔在测定眼压(IOP)方面的可靠性。患者被随机分为四组,所有压平操作均由同一名研究人员完成。最初使用的眼压笔(眼压笔#1)在用于77只眼睛后无法校准。然后获得并测试了眼压笔#2。每只眼压笔与Goldmann测量值均显示出高度相关性(双眼r均大于或等于0.86)。眼压笔检测眼压在21 mmHg或更高(由Goldmann仪器测量)的总体敏感度,右眼为62.1%,左眼为72.4%,相应的特异度,右眼为92.6%,左眼为97.1%。鉴于眼压笔#2的敏感度显著提高,眼压笔似乎适用于将21 mmHg或以上的眼压视为异常的筛查项目;然而,在较高眼压(大于或等于30 mmHg)时,眼压笔往往低估Goldmann眼压值。在低眼压(小于或等于9 mmHg)时,眼压笔往往高估眼压值。也许对该仪器进行进一步改进后,它可用于青光眼患者的临床诊断和管理。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验