患者对可移动血压监测设备的偏好:腕式还是臂式?

Patient preferences for ambulatory blood pressure monitoring devices: Wrist-type or arm-type?

机构信息

Shenzhen Baoan Women's and Children's Hospital, Jinan University, Shenzhen, Guangdong Province, China.

School of Health Sciences, Caritas Institute of Higher Education, Tseung Kwan O, Hong Kong.

出版信息

PLoS One. 2021 Aug 9;16(8):e0255871. doi: 10.1371/journal.pone.0255871. eCollection 2021.

Abstract

BACKGROUND

Ambulatory blood pressure monitoring (ABPM) is important in evaluating average 24-hour blood pressure (BP) levels, circadian rhythm, sleeping BP and BP variability but many patients are reluctant to use standard ABPM devices.

METHODS

We compared two validated ABPM devices, the BPro tonometric wrist monitor and the A&D TM-2430 oscillometric upper arm monitor, for agreement of recordings and acceptability in 37 hypertensive patients (aged 55±9 years).

RESULTS

Successful BP measurements were less frequent with the wrist-type than the arm-type device during the sleeping (66.3% vs. 92.9%, P <0.01) and awake periods (56.2% vs. 86.5%, P <0.01). Comparable paired readings showed no significant difference in systolic BP but diastolic BP (DBP) values were higher with the wrist compared to the arm monitor (24-hour 89±13 vs. 85±14 mmHg, P <0.01) with similar differences awake and sleeping. Bland-Altman analysis showed some large discrepancies between individual arm and wrist monitor measurements. More patients found the wrist monitor more comfortable to use than the arm monitor.

CONCLUSIONS

Despite the difference in individual BP measurements and the systematic overestimation of DBP values with the BPro device, wrist monitors with good patient acceptability may be useful to facilitate ABPM in some patients to provide additional information about cardiovascular risk and response to antihypertensive therapies.

摘要

背景

动态血压监测(ABPM)在评估平均 24 小时血压(BP)水平、昼夜节律、睡眠时血压和血压变异性方面非常重要,但许多患者不愿意使用标准的 ABPM 设备。

方法

我们比较了两种经过验证的 ABPM 设备,即 BPro 压力式腕式监测仪和 A&D TM-2430 振荡式上臂监测仪,以评估 37 例高血压患者(年龄 55±9 岁)的记录一致性和可接受性。

结果

与臂式设备相比,腕式设备在睡眠(66.3% vs. 92.9%,P<0.01)和清醒(56.2% vs. 86.5%,P<0.01)期间成功测量血压的频率较低。可比配对读数显示收缩压无显著差异,但舒张压(DBP)值腕式监测器高于臂式监测器(24 小时 89±13 与 85±14mmHg,P<0.01),清醒和睡眠时差异相似。Bland-Altman 分析显示臂式和腕式监测器之间存在一些较大的个体测量差异。更多的患者认为腕式监测器比臂式监测器更舒适。

结论

尽管 BPro 设备在个体 BP 测量值上存在差异,并且 DBP 值存在系统高估,但具有良好患者可接受性的腕式监测器可能有助于在某些患者中进行 ABPM,以提供有关心血管风险和抗高血压治疗反应的额外信息。