Orbit. 2022 Oct;41(5):591-597. doi: 10.1080/01676830.2021.1980894. Epub 2021 Sep 19.
This study introduces a novel exophthalmometry method utilizing a mobile platform to obtain calibrated quantitative measurements of globe position and compares its reliability to Hertel exophthalmometry.
A prospective, comparative study included 50 patients (99 eyes) for a total of 594 mobile measurements. Healthy individuals from the community in Augusta, Georgia, and individuals from a routine oculoplastics clinic at Mayo Clinic in Rochester, Minnesota, were included to represent variety within the population. The main outcome was the inter-modality correlation and agreement between Hertel and mobile exophthalmometry, and the intra-observer and inter-observer reliability with repeated mobile measurements.
There was no significant difference in the measurements obtained by Hertel exophthalmometry and mobile exophthalmometry. There was a strong linear correlation between Hertel and mobile exophthalmometry with a Pearson Correlation Coefficient of 0.910 and 0.888 for the right and left eyes, respectively ( < .001, 2-tailed). Bland-Altman plot analysis showed excellent agreement between the two modalities. The mobile platform demonstrated high intra-observer reliability with a Cronbach's alpha of 0.992 and 0.985 for the right and left eyes. An intraclass correlation coefficient of 0.992 (95% CI: 0.987-0.995) for the right eye and 0.986 (95% CI: 0.978-0.991) for the left eye demonstrated excellent reliability between observers.
Mobile exophthalmometry may be a promising tool for obtaining calibrated quantitative measurements of globe position for situations in which Hertel exophthalmometry is not available. The strong correlation and excellent agreement between Hertel and mobile measurement suggest that mobile exophthalmometry can yield reliable and accurate measurements.
本研究介绍了一种利用移动平台获取眼球位置校准定量测量的新型眼球突出度测量方法,并将其与 Hertel 眼球突出度测量法的可靠性进行比较。
这是一项前瞻性、对照研究,共纳入 50 例患者(99 只眼),共进行了 594 次移动测量。参与者包括佐治亚州奥古斯塔社区的健康个体和明尼苏达州罗切斯特梅奥诊所常规整形眼科诊所的个体,以代表人群中的多样性。主要结局是 Hertel 和移动眼球突出度测量之间的模态间相关性和一致性,以及重复移动测量时观察者内和观察者间的可靠性。
Hertel 眼球突出度测量和移动眼球突出度测量的测量值无显著差异。Hertel 和移动眼球突出度测量之间存在很强的线性相关性,右眼和左眼的 Pearson 相关系数分别为 0.910 和 0.888(<0.001,双侧)。Bland-Altman 图分析显示两种方法具有良好的一致性。移动平台具有很高的观察者内可靠性,右眼和左眼的 Cronbach's alpha 分别为 0.992 和 0.985。右眼的组内相关系数为 0.992(95%置信区间:0.987-0.995),左眼为 0.986(95%置信区间:0.978-0.991),表明观察者间具有极好的可靠性。
在无法进行 Hertel 眼球突出度测量的情况下,移动眼球突出度测量可能是一种很有前途的获取眼球位置校准定量测量的工具。Hertel 和移动测量之间的强相关性和极好的一致性表明,移动眼球突出度测量可以提供可靠和准确的测量结果。