Centre for Medical and Radiation Physics, University of Wollongong, Wollongong, Australia.
Ingham Institute for Applied Medical Research, Liverpool, Australia.
J Appl Clin Med Phys. 2021 Nov;22(11):143-150. doi: 10.1002/acm2.13432. Epub 2021 Sep 25.
To determine baseline accuracy and reproducibility of T and T relaxation times over 12 months on a dedicated radiotherapy MRI scanner.
An International Society of Magnetic Resonance in Medicine/National Institute of Standards and Technology (ISMRM/NIST) System Phantom was scanned monthly on a 3T MRI scanner for 1 year. T was measured using inversion recovery (T -IR) and variable flip angle (T -VFA) sequences and T was measured using a multi-echo spin echo (T -SE) sequence. For each vial in the phantom, accuracy errors (%bias) were determined by the relative differences in measured T and T times compared to reference values. Reproducibility was measured by the coefficient of variation (CV) of T and T measurements across monthly scans. Accuracy and reproducibility were mainly assessed on vials with relaxation times expected to be in physiological ranges at 3T.
A strong linear correlation between measured and reference relaxation times was found for all sequences tested (R > 0.997). Baseline bias (and CV[%]) for T -IR, T -VFA and T -SE sequences were +2.0% (2.1), +6.5% (4.2), and +8.5% (1.9), respectively.
The accuracy and reproducibility of T and T on the scanner were considered sufficient for the sequences tested. No longitudinal trends of variation were deduced, suggesting less frequent measurements are required following the establishment of baselines.
在专用放射治疗 MRI 扫描仪上确定 T1 和 T2 弛豫时间在 12 个月内的基线准确性和可重复性。
使用国际磁共振医学学会/美国国家标准与技术研究院(ISMRM/NIST)系统模型,在 3T MRI 扫描仪上每月扫描一次,共扫描 1 年。使用反转恢复(T1-IR)和可变翻转角(T1-VFA)序列测量 T1,使用多回波自旋回波(T1-SE)序列测量 T2。对于模型中的每个小瓶,通过测量 T1 和 T2 时间与参考值的相对差异来确定准确性误差(%偏差)。通过每月扫描的 T1 和 T2 测量值的变异系数(CV)来衡量可重复性。主要在预期在 3T 下处于生理范围内的弛豫时间的小瓶上评估准确性和可重复性。
所有测试序列均发现测量弛豫时间与参考弛豫时间之间存在很强的线性相关性(R > 0.997)。T1-IR、T1-VFA 和 T1-SE 序列的基线偏差(和 CV[%])分别为+2.0%(2.1)、+6.5%(4.2)和+8.5%(1.9)。
在所测试的序列中,扫描仪上 T1 和 T2 的准确性和可重复性被认为是足够的。没有推断出随时间变化的纵向趋势,这表明在建立基线后,需要较少的频繁测量。