Naughton Piers J, Joyce Mary, Mac Giolla Eain Marc, O'Sullivan Andrew, MacLoughlin Ronan
School of Chemistry, National University of Ireland, H91 CF50 Galway, Ireland.
Research and Development, Science and Emerging Technologies, Aerogen Limited, Galway Business Park, H91 HE94 Galway, Ireland.
Pharmaceutics. 2021 Sep 28;13(10):1574. doi: 10.3390/pharmaceutics13101574.
Drug delivery devices used for aerosol therapy during mechanical ventilation to ease the symptoms of respiratory diseases provide beneficial treatment but can also pose challenges. Reflecting the significant changes in global guidance around aerosol usage and lung-protective ventilation strategies, seen in response to the COVID-19 pandemic, for the first time, we describe the drug delivery performance of commonly used devices under these conditions. Here, vibrating mesh nebuliser (VMN), jet nebuliser (JN) and pressurised metered-dose inhaler (pMDI) performance was assessed during simulated adult mechanical ventilation. Both standard test breathing patterns and those representatives of low tidal volume (LTV) ventilation with concurrent active and passive humidification were investigated. Drug delivery using a VMN was significantly greater than that with a JN and pMDI for both standard and LTV ventilation. Humidification type did not affect the delivered dose across all device types for standard ventilation. Significant variability in the pMDI dosing was evident, depending on the timing of actuation and the adapter type used. pMDI actuation synchronised with inspiration resulted in a higher delivered drug dose. The type of adapter used for pMDI actuation influenced drug delivery, with the highest dose observed using the CombiHaler.
用于机械通气期间雾化治疗以缓解呼吸道疾病症状的给药装置虽能提供有益治疗,但也可能带来挑战。鉴于在应对新冠疫情过程中全球关于雾化使用和肺保护性通气策略的指导发生了重大变化,我们首次描述了常用装置在这些条件下的给药性能。在此,我们评估了模拟成人机械通气期间振动网式雾化器(VMN)、喷射式雾化器(JN)和压力定量吸入器(pMDI)的性能。研究了标准测试呼吸模式以及那些具有同时主动和被动加湿功能的低潮气量(LTV)通气模式。对于标准通气和LTV通气,使用VMN的给药量均显著高于使用JN和pMDI的给药量。对于标准通气,加湿类型并未影响所有装置类型的给药剂量。pMDI给药存在显著变异性,这取决于启动时间和所使用的适配器类型。pMDI与吸气同步启动可导致更高的给药剂量。用于pMDI启动的适配器类型会影响药物递送,使用组合吸入器(CombiHaler)时观察到的剂量最高。