Refractive outcomes of table-mounted and hand-held auto-refractometers in children: an observational cross-sectional study.

机构信息

Department of Ophthalmology, Mugla Sıtkı Koçman University Medical School, 48300, Mugla, Turkey.

出版信息

BMC Ophthalmol. 2021 Dec 9;21(1):424. doi: 10.1186/s12886-021-02199-5.

Abstract

BACKGROUND

To compare the refractive results of hand-held and table-mounted autorefractors.

METHODS

We designed this study as an observational, cross-sectional study. We compared the mean spheric and cylinder power, spherical equivalent, Jackson cross-cylinder values, determined the limits of agreement (LoA), and evaluated the reliability of two autorefractors.

RESULTS

We evaluated 256 eyes of 256 pediatric patients (mean age, 9.12 ± 2.26 years; range, 5-16 years). 49% of the patients were female, and 51% were male. The Nidek HandyRef-K autorefractor measured relatively more astigmatism (P < 0.001) and less hyperopia (P = 0.024). The mean differences and 95% LoA were 0.06 D ± 0.47 D (- 0.82 D to 0.98 D) in spherical power, 0.08 D ± 0.28 D (- 0.47 D to 0.64 D) in cylindrical power, 0.11 D ± 0.47 D (- 0.81 D to 1.01 D) in spherical equivalent, 0.02 D ± 0.36 D (- 0.73 D to 0.69 D) in Jackson cross-cylinder power at 0°, 0.005 D ± 0.54 D (- 1.07 D to 1.06 D) in Jackson cross-cylinder power at 45° We found the difference within 0.50 D in 244 (95%) eyes for spherical power, in 245 (96%) eyes for cylindrical power, 228 (89%) eyes for spherical equivalent, 224 (87%) eyes for Jackson cross-cylinder power at 0°, 213 (83%) eyes for Jackson cross-cylinder power at 45° When comparing devices, there were strong correlations for spherical power (Spearman's rho = 0.99, P < 0.001), cylindrical power (Spearman's rho = 0.88, P < 0.001), and spherical equivalent (Spearman's rho = 0.98, P < 0.001).

CONCLUSION

Two autorefractors showed clinically applicable agreement limits; excellent reliability for spherical power and spherical equivalent and good reliability for cylindrical power; high positive percent agreement for spherical and cylindrical power, spherical equivalent, Jackson cross-cylinder power at 0°and 45°. These results showed that both devices might be used interchangeably for screening of refractive error in children.

摘要

背景

比较手持式和台式自动验光仪的屈光结果。

方法

我们将这项研究设计为一项观察性、横断面研究。我们比较了平均球镜和柱镜力、等效球镜、Jackson 交叉圆柱镜值,确定了(差异的)一致性界限(LoA),并评估了两种自动验光仪的可靠性。

结果

我们评估了 256 例儿科患者(平均年龄 9.12±2.26 岁;范围为 5-16 岁)的 256 只眼。49%的患者为女性,51%为男性。尼德克 HandyRef-K 自动验光仪测量的散光(P<0.001)相对较多,远视(P=0.024)较少。球镜力的平均差异和 95%LoA 为 0.06 D±0.47 D(-0.82 D 至 0.98 D),柱镜力为 0.08 D±0.28 D(-0.47 D 至 0.64 D),等效球镜为 0.11 D±0.47 D(-0.81 D 至 1.01 D),Jackson 交叉圆柱镜力在 0°时为 0.02 D±0.36 D(-0.73 D 至 0.69 D),Jackson 交叉圆柱镜力在 45°时为 0.005 D±0.54 D(-1.07 D 至 1.06 D)。我们发现,在 244 只(95%)眼的球镜力、245 只(96%)眼的柱镜力、228 只(89%)眼的等效球镜力、224 只(87%)眼的 0°Jackson 交叉圆柱镜力、213 只(83%)眼的 45°Jackson 交叉圆柱镜力中,差异在 0.50 D 以内。当比较设备时,球镜力(Spearman's rho=0.99,P<0.001)、柱镜力(Spearman's rho=0.88,P<0.001)和等效球镜力(Spearman's rho=0.98,P<0.001)具有很强的相关性。

结论

两种自动验光仪的一致性界限具有临床应用价值;球镜力和等效球镜的可靠性极佳,柱镜力的可靠性良好;球镜和柱镜力、等效球镜、0°和 45°Jackson 交叉圆柱镜力的高度正百分比一致。这些结果表明,这两种设备可能可以互换用于儿童屈光不正的筛查。