Sayed Khulood Muhammad, Alsmman Alahmady Hammad, Mostafa Engy Mohammed
The Department of Ophthalmology, Sohag Faculty of Medicine, Sohag University, Sohag, Egypt.
Clin Ophthalmol. 2021 Apr 1;15:1391-1401. doi: 10.2147/OPTH.S263667. eCollection 2021.
To compare the agreement between commercially available table mounted and a hand-held autorefractors and their agreement with subjective refraction. The effect of different body position with the handheld autorefractometer was also evaluated.
A prospective study was performed on 253 healthy eyes. Refraction was acquired by a table-mounted Huvitz and hand-held Nidek autorefractometer, subjective refraction was acquired. Refractive errors were compared in terms of spherical equivalent (SE), cylinder power, and the J0 and J45. The level of agreement was evaluated by Bland-Altman plots.
There was a significant difference in SE measurements between both devices and between them and subjective refraction (P=0.00). The Huvitz SE readings tended to be less myopic. However, limits of agreement (LOA) for SE were narrowest for Nidek sitting vs supine followed by Huvitz vs subjective SE refraction. The LOA for SE for Nidek sitting vs subjective SE were of wider range. For cylinder values, LOA were similar for all devices and positions and between them and subjective cylinder refraction.
Table mounted Huvitz and Nidek portable autorefractor cannot be used interchangeably in clinical practice except for estimation of the cylinder power. No difference in refraction between sitting and supine positions for portable Nidek autorefractor but with caution in cylinder axis. High agreement was achieved between subjective refraction and Huvitz readings but not with Nidek hand-held autorefractor. A highly reliable spectacle prescription could be done based on Huvitz readings. Both devices and positions could be used interchangeably in estimation of K-readings.
比较市售台式和手持式自动验光仪之间的一致性,以及它们与主观验光的一致性。还评估了手持式自动验光仪在不同体位下的效果。
对253只健康眼睛进行前瞻性研究。使用台式Huvitz自动验光仪和手持式尼德克自动验光仪获取验光结果,并进行主观验光。比较球镜等效度(SE)、柱镜度数以及J0和J45方向上的屈光不正。通过Bland-Altman图评估一致性水平。
两种仪器之间以及它们与主观验光在SE测量上存在显著差异(P = 0.00)。Huvitz的SE读数近视程度往往较低。然而,尼德克坐姿与仰卧位时SE的一致性界限(LOA)最窄,其次是Huvitz与主观SE验光。尼德克坐姿与主观SE的SE的LOA范围更广。对于柱镜值,所有仪器和体位之间以及它们与主观柱镜验光之间的LOA相似。
在临床实践中,除了估计柱镜度数外,台式Huvitz和尼德克便携式自动验光仪不能互换使用。便携式尼德克自动验光仪在坐姿和仰卧位时的验光结果无差异,但在柱镜轴位测量时需谨慎。主观验光与Huvitz读数之间一致性较高,但与尼德克手持式自动验光仪不一致。基于Huvitz读数可以开出高度可靠的眼镜处方。在估计角膜曲率读数时,两种仪器和体位可以互换使用。