心电图门控 F-FDG 小动物 PET 的帧数对 LV 收缩和舒张功能参数有显著影响。

The Number of Frames on ECG-Gated F-FDG Small Animal PET Has a Significant Impact on LV Systolic and Diastolic Functional Parameters.

机构信息

Department of Nuclear Medicine, University Hospital Würzburg, Würzburg, Germany.

Comprehensive Heart Failure Center, University Hospital Würzburg, Würzburg, Germany.

出版信息

Mol Imaging. 2021 Dec 6;2021:4629459. doi: 10.1155/2021/4629459. eCollection 2021.

Abstract

OBJECTIVES

This study is aimed at investigating the impact of frame numbers in preclinical electrocardiogram- (ECG-) gated F-fluorodeoxyglucose (F-FDG) positron emission tomography (PET) on systolic and diastolic left ventricular (LV) parameters in rats.

METHODS

F-FDG PET imaging using a dedicated small animal PET system with list mode data acquisition and continuous ECG recording was performed in diabetic and control rats. The list-mode data was sorted and reconstructed with different numbers of frames (4, 8, 12, and 16) per cardiac cycle into tomographic images. Using an automatic ventricular edge detection software, left ventricular (LV) functional parameters, including ejection fraction (EF), end-diastolic (EDV), and end-systolic volume (ESV), were calculated. Diastolic variables (time to peak filling (TPF), first third mean filling rate (1/3 FR), and peak filling rate (PFR)) were also assessed.

RESULTS

Significant differences in multiple parameters were observed among the reconstructions with different frames per cardiac cycle. EDV significantly increased by numbers of frames (353.8 ± 57.7 l, 380.8 ± 57.2 l, 398.0 ± 63.1 l, and 444.8 ± 75.3 l at 4, 8, 12, and 16 frames, respectively; < 0.0001 vs. 16 frames), while systolic (EF) and diastolic (TPF, 1/3 FR and PFR) parameters were not significantly different between 12 and 16 frames. In addition, significant differences between diabetic and control animals in 1/3 FR and PFR in 16 frames per cardiac cycle were observed ( < 0.005), but not for 4, 8, and 12 frames.

CONCLUSIONS

Using ECG-gated PET in rats, measurements of cardiac function are significantly affected by the frames per cardiac cycle. Therefore, if you are going to compare those functional parameters, a consistent number of frames should be used.

摘要

目的

本研究旨在探讨在临床前心电图门控 F-氟代脱氧葡萄糖(F-FDG)正电子发射断层扫描(PET)中,帧数量对大鼠收缩和舒张左心室(LV)参数的影响。

方法

使用带有连续心电图记录的专用小动物 PET 系统进行 F-FDG PET 成像。使用列表模式数据采集和连续心电图记录。使用自动心室边缘检测软件,计算左心室(LV)功能参数,包括射血分数(EF)、舒张末期(EDV)和收缩末期容积(ESV)。还评估了舒张变量(峰值充盈时间(TPF)、第一三分之一充盈率(1/3 FR)和峰值充盈率(PFR))。

结果

在不同的帧数量重建中,多个参数存在显著差异。EDV 随帧数量增加而显著增加(分别为 353.8±57.7 l、380.8±57.2 l、398.0±63.1 l 和 444.8±75.3 l; <0.0001 与 16 帧相比),而收缩(EF)和舒张(TPF、1/3 FR 和 PFR)参数在 12 和 16 帧之间没有显著差异。此外,在 16 帧/心动周期中,糖尿病和对照组动物之间的 1/3 FR 和 PFR 存在显著差异( <0.005),但在 4、8 和 12 帧时没有差异。

结论

在大鼠中使用心电图门控 PET,心脏功能的测量受到心动周期中帧数量的显著影响。因此,如果要比较这些功能参数,应使用一致的帧数。