Rotational tolerances of a titanium abutment in the as-received condition and after screw tightening in a conical implant connection.

作者信息

Prisco Rosario, Troiano Giuseppe, Laino Luigi, Zhurakivska Khrystyna

机构信息

Department of Clinical and Experimental Medicine, University of Foggia, Foggia, Italy.

Multidisciplinary Department of Medical-Surgical and Dental Specialties, University of Campania - Luigi Vanvitelli, Naples, Italy.

出版信息

J Adv Prosthodont. 2021 Dec;13(6):343-350. doi: 10.4047/jap.2021.13.6.343. Epub 2021 Dec 22.

Abstract

PURPOSE

The success of an implant-prosthetic rehabilitation is influenced by good implant health and an excellent implant-prosthetic coupling. The stability of implant-prosthetic connection is influenced by the rotational tolerance between anti-rotational features on the implant and those on the prosthetic component. The aim of this study is to investigate the rotational tolerance of a conical connection implant system and its titanium abutment counterpart, in various conditions.

MATERIAL AND METHODS

10 preparable titanium abutments, having zero-degree angulation (MegaGen, Daegu, Korea) with an internal 5-degree conical connection, and 10 implants (MegaGen, Daegu, Korea) were used. Rotational tolerance between the connection of implant and titanium abutments was measured through the use of a tridimensional optics measuring system (Quick Scope QS250Z, Mitutoyo, Kawasaki, Japan) in the as-received condition (Time 0), after securing with a titanium screw tightening at 35 Ncm (Time 1), after tightening 4 times at 35 Ncm (Time 2), after tightening one more time at 45 Ncm (Time 3), and after tightening another 4 times at 45 Ncm (Time 4).

RESULTS

The group "Time 0" had the lowest values of rotational freedom (0.22 ± 0.76 degrees), followed by the group Time 1 (0.46 ± 0.83 degrees), the group Time 2 (1.01 ± 0.20 degrees), the group Time 3 (1.30 ± 0.85 degrees), and the group Time 4 (1.49 ± 0.17 degrees).

CONCLUSION

The rotational tolerance of a conical connection is low in the "as received" condition but increases with repetitive tightening and with application of a torque greater than 35 Ncm.

摘要

目的

种植体修复的成功受良好的种植体健康状况和出色的种植体与修复体连接的影响。种植体与修复体连接的稳定性受种植体上的抗旋转特征与修复体部件上的抗旋转特征之间的旋转公差影响。本研究的目的是在各种条件下研究锥形连接种植体系统及其钛基台对应物的旋转公差。

材料与方法

使用10个可预备的钛基台(韩国大邱MegaGen公司),其具有零度角度,内部为5度锥形连接,以及10个种植体(韩国大邱MegaGen公司)。种植体与钛基台连接之间的旋转公差通过三维光学测量系统(日本川崎三丰公司的Quick Scope QS250Z)在接收状态(时间0)、以35 Ncm的扭矩拧紧钛螺钉后(时间1)、以35 Ncm拧紧4次后(时间2)、以45 Ncm再拧紧一次后(时间3)以及以45 Ncm再拧紧4次后(时间4)进行测量。

结果

“时间0”组的旋转自由度值最低(0.22±0.76度),其次是时间1组(0.46±0.83度)、时间2组(1.01±0.20度)、时间3组(1.30±0.85度)和时间4组(1.49±0.17度)。

结论

锥形连接在“接收时”的旋转公差较低,但随着反复拧紧以及施加大于35 Ncm的扭矩而增加。