Blakeman Thomas, Fowler John-Michael, Salvator Ann, Rodriquez Dario
Mil Med. 2022 Feb 4. doi: 10.1093/milmed/usab561.
Military transport of critically ill/injured patients requires judicious use of resources. Maintaining oxygen (O2) supplies for mechanically ventilated is crucial. O2 cylinders are difficult to transport due to the size and weight and add the risk of fire in an aircraft. The proposed solution is the use of a portable oxygen concentrator (POC) to supply O2 for mechanical ventilation. As long as power is available, a POC can provide an endless supply of O2. Anecdotal evidence suggests that as little as 3 L/min of O2 could manage as many as 2/3 of the mechanically ventilated military aeromedical transport patients.
We evaluated two each of the AutoMedx SAVe II, Hamilton T1, Zoll 731, and Ventec VOCSN portable ventilators over a range of settings paired with 1 and 2 Caire SAROS POCs at ground level and simulated altitudes of 8,000 feet, 16,000 feet, and 22,000 feet. The Ventec VOCSN has the capability of utilizing an internal O2 concentrator that uses pulsed dose technology, which was also evaluated. Each ventilator was attached to a Michigan Instruments Training Test Lung. Output from the POC was bled into each ventilator via the mechanism provided with each device. A Fleisch pneumotach was used to measure delivered tidal volume (VT), and a fast-response O2 analyzer was used to measure FiO2 within the simulated lung. Ventilator parameters and FiO2 were continuously measured and recorded at each altitude. One-way analysis of variance was used to determine statistically significant differences (P < .05) in FiO2 between ventilators and among the same ventilator model at each testing condition.
Delivered FiO2 varied widely between ventilator models and between devices of the same model with some testing conditions. Differences in FiO2 between ventilators at a majority (98.5%) of testing conditions were statistically significant (P < .05) but not all were clinically important. The Zoll 731 delivered the highest and most consistent FiO2 over all ventilator/POC settings at all altitudes. Differences in FiO2 at a given ventilator/POC setting from ground level to 22,000 feet were not clinically important (<5%) with this device. The VOCSN utilizing the integrated internal O2 concentrator delivered the lowest FiO2 across all ventilator/POC settings and altitudes. Due to the inability of the SAVe II to operate at the minute ventilation and positive end expiratory pressure (PEEP) settings required by the testing protocol, the device was only tested at one ventilator setting. The Hamilton T1 failed to operate appropriately at the highest VT/PEEP setting at 16,000 feet and all but one ventilator setting at 22,000 feet. The delivered FiO2 was not included in the analysis for those ventilator settings. The highest delivered FiO2 was 0.85 ± 0.05 at the 250 mL VT setting using 2 POCs (P < .0001) at ground level with the Zoll 731.
Oxygen delivery utilizing POCs is dependent upon multiple factors including ventilator operating characteristics, ventilator settings, altitude, and the use of pulsed dose or continuous flow O2. Careful patient selection would be paramount to provide safe mechanical ventilation using this method of O2 delivery.
危重伤病员的军事运输需要合理使用资源。为机械通气患者维持氧气(O₂)供应至关重要。氧气瓶由于尺寸和重量问题难以运输,且会增加飞机起火风险。提议的解决方案是使用便携式制氧机(POC)为机械通气提供氧气。只要有电源,POC就能提供源源不断的氧气。有传闻证据表明,低至每分钟3升的氧气就能满足多达三分之二的机械通气军事航空医疗运输患者的需求。
我们在一系列设置条件下,分别对两台AutoMedx SAVe II、两台Hamilton T1、两台Zoll 731和两台Ventec VOCSN便携式呼吸机进行了评估,这些呼吸机分别与1台和2台Caire SAROS POC在地面以及模拟海拔8000英尺、16000英尺和22000英尺的高度配对使用。Ventec VOCSN具备使用采用脉冲剂量技术的内置制氧机的能力,对此也进行了评估。每台呼吸机都连接到一台密歇根仪器训练测试肺上。POC的输出通过各设备配备的装置注入每台呼吸机。使用费利氏呼吸流速计测量呼出潮气量(VT),并使用快速响应氧气分析仪测量模拟肺内的吸入氧浓度(FiO₂)。在每个海拔高度持续测量并记录呼吸机参数和FiO₂。采用单因素方差分析来确定各测试条件下不同呼吸机之间以及同一型号呼吸机不同设备之间FiO₂的统计学显著差异(P < 0.05)。
在某些测试条件下,不同呼吸机型号之间以及同一型号不同设备之间的吸入氧浓度差异很大。在大多数(98.5%)测试条件下,不同呼吸机之间的FiO₂差异具有统计学意义(P < 0.05),但并非所有差异都具有临床重要性。在所有海拔高度的所有呼吸机/POC设置中,Zoll 731提供的FiO₂最高且最稳定。对于该设备,从地面到22000英尺,在给定的呼吸机/POC设置下FiO₂的差异在临床上并不重要(<5%)。在所有呼吸机/POC设置和海拔高度下,使用集成内置制氧机的VOCSN提供的FiO₂最低。由于SAVe II无法在测试方案要求的分钟通气量和呼气末正压(PEEP)设置下运行,该设备仅在一种呼吸机设置下进行了测试。Hamilton T1在16000英尺的最高VT/PEEP设置下以及在22000英尺除一种呼吸机设置外的所有设置下均无法正常运行。对于那些呼吸机设置,未将所提供的FiO₂纳入分析。在地面使用2台POC且VT设置为250毫升时,Zoll 731提供的最高FiO₂为0.85 ± 0.05(P < 0.0001)。
使用POC进行氧气输送取决于多个因素,包括呼吸机的运行特性、呼吸机设置、海拔高度以及脉冲剂量或连续流氧气的使用。谨慎选择患者对于使用这种氧气输送方法提供安全的机械通气至关重要。