Fluid warming, recommended for fluid rates of > 500 ml h, is an integral part of patient temperature management strategies. Fluid warming devices using an uncoated aluminium containing heating element have been reported to liberate aluminium resulting in critical aluminium concentrations in heated fluids. We investigated saline solution (0.9%), artificially spiked with organic acids to determine the influence of fluid composition on aluminium release using the uncoated enFlow® device. Additionally, the Level1® as a high volume fluid warming device and the ThermoSens® device were investigated with artificial spiked fluid at high risk for aluminum release and a clinically used crystalloid solution.
Saline solution spiked with lactate more than acetate, especially at a non neutral pH, led to high aluminium release. Next to the enFlow® device, aluminium release was observed for the Level1® device, but not for the coated ThermoSens®-device.
Uncoated aluminium containing fluid warming devices lead to potentially toxic levels of aluminium in heated fluids, especially in fluids with non-neutral pH containing organic acids and their salts like balanced electrolyte solutions.
对于流速大于 500ml/h 的液体,建议进行液体加温,这是患者体温管理策略的一个组成部分。含有未涂覆铝的加热元件的液体加温设备已被报道会释放铝,导致加热液体中出现临界浓度的铝。我们使用未涂覆的 enFlow®设备研究了生理盐水(0.9%),并用有机酸人为地对其进行了人工加标,以确定液体成分对铝释放的影响。此外,我们还使用人工加标高风险的铝释放液体和一种临床使用的晶体溶液对 Level1®高容量液体加温设备和 ThermoSens®设备进行了研究。
与乙酸相比,乳酸盐在生理盐水(0.9%)中的加标导致了更高的铝释放。除了 enFlow®设备之外,铝释放还发生在 Level1®设备中,但在涂覆的 ThermoSens®-设备中没有观察到。
含有未涂覆铝的液体加温设备会导致加热液体中出现潜在毒性水平的铝,特别是在 pH 值非中性的含有有机酸及其盐(如平衡电解质溶液)的液体中。