Jones C, MacPherson D W, Stevens D L
J Clin Microbiol. 1986 Jan;23(1):160-2. doi: 10.1128/jcm.23.1.160-162.1986.
The Chemstrip LN (Boehringer Mannheim Biochemicals, Indianapolis, Ind.), designed to detect pyuria and bacteriuria, was compared with culture of 1,020 unselected, consecutive midstream urine specimens and evaluated on its ability to predict colony counts at three levels. At the level of greater than or equal to 10(5) CFU/ml, the combined test (detection of leukocyte esterase and nitrite) had sensitivity of 82.3%, specificity of 67.9%, positive predictive value of 41.3%, and negative predictive value of 93.3% at prevalence rate of 21.6%. The test would have rejected 9.4% of the specimens with significant bacteriuria if the Chemstrip alone had been used.
旨在检测脓尿和菌尿的Chemstrip LN(德国勃林格殷格翰生物化学制品公司,印第安纳波利斯,印第安纳州),与1020份未经挑选的连续中段尿标本的培养结果进行了比较,并评估了其在三个水平上预测菌落计数的能力。在菌落数大于或等于10⁵CFU/ml的水平,联合检测(白细胞酯酶和亚硝酸盐检测)在患病率为21.6%时,敏感性为82.3%,特异性为67.9%,阳性预测值为41.3%,阴性预测值为93.3%。如果仅使用Chemstrip检测,该检测会排除9.4%有显著菌尿的标本。