Chen Shihao, Zhang Qiaoyue, Savini Giacomo, Zhang Shuangzhe, Huang Xiaomin, Yu Jinjin, Wang Yirang, Ning Rui, Huang Jinhai, Tu Ruixue
Eye Hospital and School of Ophthalmology and Optometry, Wenzhou Medical University; State Key Laboratory of Optometry, Ophthalmology and Vision Science, Wenzhou, China.
Department of Ophthalmology, Air Force Medical Center, Beijing, China.
Front Med (Lausanne). 2022 Feb 10;8:814519. doi: 10.3389/fmed.2021.814519. eCollection 2021.
To evaluate measurement precision and to compare the Pentacam AXL (Oculus Optikgeräte, Wetzlar, German), a new optical biometer based on Scheimpflug imaging and partial coherence interferometry (PCI) with that of the OA-2000 biometer (Tomey, Nagoya, Japan), which combines swept-source optical coherence tomography (SS-OCT) and Placido-disk topography.
Axial length (AL), central corneal thickness (CCT), anterior chamber depth (ACD), aqueous depth (AQD), mean keratometry (Km), astigmatism vectors J0, J45, and corneal diameter (CD) were measured in triplicate by two technical operators. Within-subject standard deviation (Sw), repeatability and reproducibility (2.77 Sw), coefficient of variation (CoV), and intraclass correlation coefficient (ICC) were used to assess the Pentacam AXL intra-observer repeatability and inter-observer reproducibility. Paired -test and Bland-Altman plots were used to determine the agreement between the two biometers.
The new optical biometer had high intra-observer repeatability [all parameters evaluated had low CoV (<0.71%) and high ICC (>0.88)]. Inter-observer reproducibility was also excellent, with high ICC (>0.95) and low CoV (<0.52%). The 95% LoA between the new biometer and OA-2000 were insignificant for most of the parameters evaluated, especially for AL. However, the measurement agreement was moderate for CCT.
Intra-observer repeatability and inter-observer reproducibility were excellent for all parameters evaluated using the new optical biometer based on Scheimpflug imaging and PCI. There was a high agreement between the two devices and hence could be clinically interchangeable for the measurement of most ocular parameters.
评估测量精度,并将基于Scheimpflug成像和部分相干干涉测量法(PCI)的新型光学生物测量仪Pentacam AXL(德国韦茨拉尔奥culus Optikgeräte公司)与OA - 2000生物测量仪(日本名古屋Tomey公司)进行比较,后者结合了扫频光学相干断层扫描(SS - OCT)和Placido盘地形图。
由两名技术操作人员对眼轴长度(AL)、中央角膜厚度(CCT)、前房深度(ACD)、房水深度(AQD)、平均角膜曲率(Km)、散光矢量J0、J45和角膜直径(CD)进行三次测量。采用受试者内标准差(Sw)、重复性和再现性(2.77Sw)、变异系数(CoV)和组内相关系数(ICC)来评估Pentacam AXL的观察者内重复性和观察者间再现性。采用配对t检验和Bland - Altman图来确定两种生物测量仪之间的一致性。
新型光学生物测量仪具有较高的观察者内重复性[所有评估参数的CoV均较低(<0.7,1%),ICC较高(>0.88)]。观察者间再现性也非常好,ICC较高(>0.95),CoV较低(<0.52%)。对于大多数评估参数,尤其是AL,新型生物测量仪与OA - 2000之间的95%一致性界限不显著。然而,CCT的测量一致性为中等。
对于基于Scheimpflug成像和PCI的新型光学生物测量仪所评估的所有参数,观察者内重复性和观察者间再现性都非常好。两种设备之间具有高度一致性,因此在测量大多数眼部参数时在临床上可相互替代。