Repeatability and Agreement of Central Corneal Thickness and Keratometry Measurements between Four Different Devices.

作者信息

Kiraly Laszlo, Stange Jana, Kunert Kathleen S, Sel Saadettin

机构信息

Augen- und Laserzentrum Leipzig, Lampestraße 1, 04107 Leipzig, Germany.

Ernst-Abbe-Hochschule Jena, Carl-Zeiß-Promenade 2, 07745 Jena, Germany.

出版信息

J Ophthalmol. 2017;2017:6181405. doi: 10.1155/2017/6181405. Epub 2017 Mar 5.

Abstract

To estimate repeatability and comparability of central corneal thickness (CCT) and keratometry measurements obtained by four different devices in healthy eyes. Fifty-five healthy eyes from 55 volunteers were enrolled in this study. CCT (IOLMaster 700, Pentacam HR, and Cirrus HD-OCT) and keratometry readings (IOLMaster 700, Pentacam HR, and iDesign) were measured. For statistical analysis, the corneal spherocylinder was converted into power vectors (J0, J45). Repeatability was assessed by intraclass correlation coefficient (ICC). Agreement of measurements between the devices was evaluated by the Bland-Altman method. The analysis of repeatability of CCT data of IOLMaster 700, Pentacam HR, and Cirrus HD-OCT showed high ICCs (range 0.995 to 0.999). The comparison of CCT measurements revealed statistically significant differences between Pentacam HR versus IOLMaster 700 ( < 0.0001) and Pentacam HR versus Cirrus HD-OCT ( < 0.0001), respectively. There was no difference in CCT measurements between IOLMaster 700 and Cirrus HD-OCT ( = 0.519). The repeatability of keratometry readings (J0 and J45) of IOLMaster 700, Pentacam HR, and iDesign was also high with ICCs ranging from 0.974 to 0.999. The Pentacam HR revealed significantly higher J0 in comparison to IOLMaster 700 ( = 0.009) and iDesign ( = 0.041); however, no significant difference was between IOLMaster 700 and iDesign ( = 0.426). Comparison of J45 showed no significant difference between IOLMaster 700, Pentacam HR, and iDesign. These results were in accordance with Bland-Altman plots. In clinical practice, the devices analyzed should not be used interchangeably due to low agreement regarding CCT as well as keratometry readings.

摘要

评估四种不同设备在健康眼睛中获得的中央角膜厚度(CCT)和角膜曲率测量值的可重复性和可比性。本研究纳入了55名志愿者的55只健康眼睛。测量了CCT(IOLMaster 700、Pentacam HR和Cirrus HD - OCT)和角膜曲率读数(IOLMaster 700、Pentacam HR和iDesign)。为进行统计分析,将角膜球柱镜转换为屈光力矢量(J0、J45)。通过组内相关系数(ICC)评估可重复性。采用Bland - Altman方法评估设备之间测量值的一致性。IOLMaster 700、Pentacam HR和Cirrus HD - OCT的CCT数据可重复性分析显示ICC值较高(范围为0.995至0.999)。CCT测量值的比较显示,Pentacam HR与IOLMaster 700之间(<0.0001)以及Pentacam HR与Cirrus HD - OCT之间(<0.0001)分别存在统计学显著差异。IOLMaster 700和Cirrus HD - OCT之间的CCT测量值无差异(=0.519)。IOLMaster 700、Pentacam HR和iDesign的角膜曲率读数(J0和J45)的可重复性也很高,ICC范围为0.974至0.999。与IOLMaster 700(=0.009)和iDesign(=0.041)相比,Pentacam HR显示出显著更高 的J0;然而,IOLMaster 700和iDesign之间无显著差异(=0.426)。J45的比较显示IOLMaster 700、Pentacam HR和iDesign之间无显著差异。这些结果与Bland - Altman图一致。在临床实践中,由于CCT以及角膜曲率读数的一致性较低,所分析的设备不应互换使用。

https://cdn.ncbi.nlm.nih.gov/pmc/blobs/edcc/5357553/d21113d5b855/JOPH2017-6181405.001.jpg

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验