LX-8000R and UriSed 2 fully automated urine analyzers comparison to manual microscopic examination.

作者信息

Evin Canpolat Erkan Revşa, Aslan Özgür

机构信息

Health Sciences University Diyarbakir Gazi Yaşargil Education and Research Hospital, Department of Medical Biochemistry, Diyarbakir, Turkey.

出版信息

J Med Biochem. 2022 Feb 2;41(1):91-99. doi: 10.5937/jomb0-31711.

Abstract

BACKGROUND

Urinalysis has an important place in evaluating kidney and urinary tract infections. Automated urine analyzers enhance productivity and turnover in laboratories and economize time and labor required for analysis. In the present study, we evaluated and compared analytic and diagnostic performance of UriSed2 with LX-8000R, which is a novel image-based automated urine sediment analyzer.

METHODS

A total of 178 urine samples sent to our laboratory were evaluated by the two urine analyzers and standard manual microscopy. Precision and comparison studies were done in accordance with CLSI guidelines.

RESULTS

Sensitivity assessment revealed similar outcomes with both UriSed2 and LX-8000R devices for erythrocyte count (RBC), whereas UriSed2 device yielded higher outcomes for leukocyte count (WBC) and epithelial cells (EPI) than LX-8000R analyzer. Specificity of UriSed2 for WBC and RBC was higher than that of LX-8000R device. According to Gamma statistics, both urine analyzers showed perfect consistency for WBC, RBC and EPI cell counts. Manuel microscopy revealed statistically significant correlation between LX-8000R and UriSed2 in terms of WBC and RBC. Manual evaluation by Bland-Altman analysis demonstrated lower WBC and RBC values and higher EPI as compared to both UriSed2 and LX-8000R devices. As the result of Passing-Bablok regression analysis, both devices were found to be inconsistent with manual microscopy.

CONCLUSIONS

We think that evaluation of automated urine analyzers will be more meaningful when they are evaluated together with urine samples and patient clinical findings in addition to comparing with manual microscopy.

摘要

背景

尿液分析在评估肾脏和尿路感染中具有重要地位。自动尿液分析仪提高了实验室的工作效率和周转速度,节省了分析所需的时间和人力。在本研究中,我们评估并比较了UriSed2与LX - 8000R(一种新型的基于图像的自动尿沉渣分析仪)的分析和诊断性能。

方法

两台尿液分析仪和标准手工显微镜检查对总共178份送至我们实验室的尿液样本进行了评估。根据CLSI指南进行了精密度和比较研究。

结果

敏感性评估显示,UriSed2和LX - 8000R设备在红细胞计数(RBC)方面结果相似,而UriSed2设备在白细胞计数(WBC)和上皮细胞(EPI)方面的结果高于LX - 8000R分析仪。UriSed2对WBC和RBC的特异性高于LX - 8000R设备。根据伽马统计,两台尿液分析仪在WBC、RBC和EPI细胞计数方面显示出完美的一致性。手工显微镜检查显示,LX - 8000R和UriSed2在WBC和RBC方面具有统计学上的显著相关性。通过Bland - Altman分析进行的手工评估表明,与UriSed2和LX - 8000R设备相比,WBC和RBC值较低,EPI值较高。通过Passing - Bablok回归分析结果发现,两台设备与手工显微镜检查不一致。

结论

我们认为,除了与手工显微镜检查进行比较外,将自动尿液分析仪与尿液样本和患者临床发现一起评估时,其评估将更有意义。