使用皮肤电极和隐形眼镜电极同步记录时视网膜电图信噪比的比较。

Comparison of ERG signal-to-noise ratios in simultaneous recordings with skin electrodes and contact lens electrodes.

作者信息

Brodie Scott E, Faruque Promie, Pincay Jorge, Sylla Mohamed, Cui Xuan, Choi Stephanie, Holopigian Karen, Greenstein Vivienne

机构信息

Department of Ophthalmology, Edward S. Harkness Eye Institute, Columbia University Vagelos College of Physicians and Surgeons, 622West 168th Street, New York, NY, 10032, USA.

SUNY Downstate Medical Center, Brooklyn, NY, USA.

出版信息

Doc Ophthalmol. 2025 Feb 4. doi: 10.1007/s10633-025-10003-8.

Abstract

PURPOSE

To compare signal-to-noise levels in ERG recordings obtained with contact lens electrodes and adhesive skin electrodes.

METHODS

23 subjects were studied. Full-field ERGs were recorded according to ISCEV standards simultaneously with ERG-jet corneal contact lens electrodes and LKC Technologies Sensor Strip adhesive skin electrodes. B-wave amplitude or peak-to-peak amplitude was used as a measure of signal strength. Noise was estimated using the " ± averaging method." Comparisons between signal strength, absolute noise levels, and signal-to-noise ratios between contact lens and skin electrodes were performed by linear regression.

RESULTS

Comparisons of signal strength for LA 3, 30-Hz, DA 0.01, and DA 3 responses, yielded regression coefficient ß values of 0.37, 0.39, 0.39, and 0.35, respectively. For the entire data set, the regression coefficient ß value was 0.36 (95% confidence limits 0.34 - 0.38). The grand average ERG noise for all ERG stimuli was 13.8 µV for contact lens electrodes and 13.0 µV for skin electrodes (not significant: p = 0.66 for paired t-test). For signal-to-noise ratios, regression ß coefficients for contact lens and adhesive skin electrodes for LA 3, 30-Hz, DA 0.01, and DA 3 stimuli were 0.25, 0.39, 0.50, and 0.36 respectively. The ß coefficient for the amalgamated data set was 0.33 (95% confidence limits 0.30- 0.36).

CONCLUSIONS

Overall ERG amplitudes obtained with skin electrodes were 1/3 those obtained with contact lens electrodes. Absolute noise levels were similar. Signal-to-noise levels with skin electrodes were 1/3 those seen with contact lens electrodes. Implications for signal-averaging in clinical applications are discussed.

摘要

目的

比较使用隐形眼镜电极和皮肤粘贴电极进行视网膜电图(ERG)记录时的信噪比。

方法

对23名受试者进行研究。根据国际临床视觉电生理学会(ISCEV)标准,使用ERG-jet角膜隐形眼镜电极和LKC Technologies Sensor Strip皮肤粘贴电极同时记录全视野ERG。以B波振幅或峰峰值振幅作为信号强度的指标。使用“±平均法”估计噪声。通过线性回归对隐形眼镜电极和皮肤电极之间的信号强度、绝对噪声水平及信噪比进行比较。

结果

对LA 3、30Hz、DA 0.01和DA 3反应的信号强度进行比较,回归系数β值分别为0.37、0.39、0.39和0.35。对于整个数据集,回归系数β值为0.36(95%置信区间0.34 - 0.38)。所有ERG刺激的总体平均ERG噪声,隐形眼镜电极组为13.8µV,皮肤电极组为13.0µV(无显著差异:配对t检验p = 0.66)。对于信噪比,LA 3、30Hz、DA 0.01和DA 3刺激下隐形眼镜电极和皮肤粘贴电极的回归β系数分别为0.25、0.39、0.50和0.36。合并数据集的β系数为0.33(95%置信区间0.30 - 0.36)。

结论

皮肤电极获得的总体ERG振幅是隐形眼镜电极获得的振幅的1/3。绝对噪声水平相似。皮肤电极的信噪比是隐形眼镜电极的1/3。文中讨论了其在临床应用中对信号平均的影响。