Reichwein B, Jungkind D, Guardiani M, Gilbert R, Prosswimmer G, Amadio P
Am J Clin Pathol. 1986 Oct;86(4):529-32. doi: 10.1093/ajcp/86.4.529.
Throat swabs from 404 patients with suspected pharyngitis were collected using duplicate swabs. Both swabs were used to inoculate 5% sheep blood agar plates, which were incubated in an anaerobic atmosphere for the isolation of Group A streptococci. The throat swabs were tested for the presence of Group A antigen using the Culturette Brand 10-Minute Group A Strep ID kit (Marion Scientific, Kansas City, MO), and the Direct Antigen Identification D.A.I. Strep A Test (Difco Laboratories, Inc., Detroit, MI). We found that 77 of the 404 specimens were culture positive for Group A streptococci. The Strep ID kit had a sensitivity of 83.7% and a specificity of 91.6%. The positive and negative predictive values were 72% and 95.6%, respectively. The D.A.I. test had a sensitivity of 80.2% and a specificity of 100%. The positive and negative predictive values were 100% and 94.5%, respectively. There was not a significant difference in the sensitivity of the two kits (P less than 0.1), but there was a significant difference in the specificity (P less than 0.01).
使用双份拭子收集了404例疑似咽炎患者的咽拭子。两份拭子均用于接种5%羊血琼脂平板,将平板置于厌氧环境中培养以分离A组链球菌。使用Culturette品牌10分钟A组链球菌鉴定试剂盒(马里恩科学公司,密苏里州堪萨斯城)和直接抗原鉴定D.A.I. A组链球菌检测试剂盒(迪夫科实验室公司,密歇根州底特律)对咽拭子进行A组抗原检测。我们发现,404份标本中有77份A组链球菌培养呈阳性。链球菌鉴定试剂盒的灵敏度为83.7%,特异性为91.6%。阳性和阴性预测值分别为72%和95.6%。D.A.I.检测的灵敏度为80.2%,特异性为100%。阳性和阴性预测值分别为100%和94.5%。两种试剂盒的灵敏度无显著差异(P小于0.1),但特异性有显著差异(P小于0.01)。