3T 肝脏扩散加权成像中超高 b 值(3000 s/mm)和最小回波时间对图像质量和 T2 穿透效应的影响:体模和临床初步研究。

Effect of an Ultrahigh b Value of 3000 s/mm and the Minimal Echo-time on Image Quality and the T2 Shine-through Effect in Diffusion-weighted Imaging of the Liver at 3T: Phantom and Clinical Pilot Studies.

机构信息

Department of Radiology, Kyorin University Hospital.

Department of Radiology, Faculty of Medicine, Kyorin University.

出版信息

Magn Reson Med Sci. 2023 Apr 1;22(2):232-240. doi: 10.2463/mrms.mp.2021-0077. Epub 2022 May 11.

Abstract

PURPOSE

To assess the effect of an ultrahigh b value of 3000 s/mm and the minimal TE of 53 ms on image quality and T2 shine-through effect in liver diffusion-weighted imaging (DWI) using a 3-Tesla MRI scanner with a peak gradient of 100 mT/m.

METHODS

At b values of 1000 and 3000 s/mm and at the minimal (44-53 ms) and routine TEs (70 ms), DWI of our original phantom and liver DWI in 10 healthy volunteers and 26 patients with 35 hepatic hemangiomas were acquired with this scanner, and the quantified SNR of the phantom and the hepatic parenchyma in the volunteers and the contrast-to-noise ratio (CNR) of the hepatic hemangiomas were calculated; two independent readers qualitatively graded the overall image quality in the volunteers and determined the presence or absence of the T2 shine-through effect related to the hemangiomas in the patients. We compared the SNR and subjective overall image quality between the minimal and routine TEs and the CNR and incidence of the T2 shine-through effect between b values of 1000 and 3000 s/mm. Inter-reader agreement was also evaluated.

RESULTS

The SNR at both b values was significantly higher, and the subjective overall image quality at a b value of 3000 s/mm was significantly better at the minimal TE than at the routine TE (P < 0.05 for all). The CNR at both TEs and the incidence of the T2 shine-through effect at the minimal TE were significantly lower at a b value of 3000 s/mm than at a b value of 1000 s/mm (P < 0.05 for all). Inter-reader agreement was excellent.

CONCLUSION

Liver DWI at the ultrahigh b value can reduce the T2 shine-through effect with improvement of image quality using the minimal TE.

摘要

目的

在具有 100 mT/m 峰值梯度的 3.0T 磁共振成像扫描仪上,评估超高质量 b 值 3000 s/mm 和最小 TE 53ms 对肝脏弥散加权成像(DWI)的图像质量和 T2 透过效应的影响。

方法

在 b 值为 1000 和 3000 s/mm 以及最小(44-53 ms)和常规 TE(70 ms)下,使用该扫描仪对我们的原始体模和 10 名健康志愿者的肝脏 DWI 以及 26 名患有 35 个肝脏血管瘤的患者的肝脏 DWI 进行采集,并计算体模和志愿者肝脏实质的量化信噪比以及肝脏血管瘤的对比噪声比(CNR);两名独立的读者对志愿者的整体图像质量进行定性分级,并确定患者中与血管瘤相关的 T2 透过效应的存在与否。我们比较了最小和常规 TE 之间的 SNR 和主观整体图像质量以及 1000 和 3000 s/mm 之间的 b 值的 CNR 和 T2 透过效应的发生率。还评估了读者间的一致性。

结果

在最小 TE 时,两种 b 值的 SNR 均显著升高,而在 3000 s/mm b 值时,主观整体图像质量也显著改善(所有 P 值均 <0.05)。在最小 TE 时,两种 TE 的 CNR 以及最小 TE 时的 T2 透过效应发生率均显著低于 1000 s/mm b 值时(所有 P 值均 <0.05)。读者间的一致性非常好。

结论

使用最小 TE,超高 b 值的肝脏 DWI 可降低 T2 透过效应,同时改善图像质量。