Rotar Raul Nicolae, Faur Andrei Bogdan, Pop Daniel, Jivanescu Anca
Department of Prosthodontics, University of Medicine and Pharmacy Victor Babes, B-dul Revolutiei 1989, No. 9, 300580 Timisoara, Romania.
TADERP Research Center, 300041 Timisoara, Romania.
Materials (Basel). 2022 Apr 22;15(9):3061. doi: 10.3390/ma15093061.
Intraoral scanners (IOS) have reached a point where their impact in the dental office cannot be denied. The distance between the tip of the IOS and the preparation may have implications on the accuracy of the digital model. The objective of this study was to evaluate the differences in accuracy between digital impressions in the scenario of different scanning distances. Twenty consecutive scans were performed at five predetermined distances: 5 mm, 10 mm, 15 mm, 20 mm and 23 mm by a single operator. The scanning distance of 10 mm displayed the best accuracy with an overall trueness value of 23.05 μm and precision value of 4.2 μm. The drawn conclusion was that increased scanning distances can decrease the accuracy of a digital impression.
口腔内扫描仪(IOS)已发展到其在牙科诊所的影响力不容小觑的程度。IOS探头尖端与预备体之间的距离可能会对数字模型的准确性产生影响。本研究的目的是评估在不同扫描距离情况下数字印模准确性的差异。由一名操作人员在五个预定距离(5毫米、10毫米、15毫米、20毫米和23毫米)下连续进行20次扫描。10毫米的扫描距离显示出最佳准确性,总体真实度值为23.05微米,精度值为4.2微米。得出的结论是,增加扫描距离会降低数字印模的准确性。