To investigate the repeatability of Anterion and compare the agreement of ocular biometric measurements and predicted intraocular lens (IOL) powers with other three optical biometers.
Flat keratometry (Kf), steep keratometry (Ks), J0 and 45 vectors, central cornea thickness (CCT), anterior chamber depth (ACD), and axial length (AL) from the Anterion, IOLMaster 700, Lenstar LS 900, and OA-2000 were recorded. The IOL powers were calculated with the Hoffer Q, Holladay 1, SRK/T, and Haigis formulas. The repeatability was evaluated using the within-subject standard deviation (Sw), repeatability coefficient (RC), coefficient of variation (COV), and intraclass correlation coefficient (ICC). Inter-device agreement between the four biometers was assessed with the 95% limits of agreement.
In total, 101 right eyes of 101 participants were enrolled. The Anterion showed good repeatability for all the included biometric parameters with all the CoV ≤ 0.30% and ICC ≥ 0.930 except for J45 with moderate repeatability (ICC was 0.849). Good agreement was found among the four devices for Kf, Ks, J0, J45, ACD, and AL. Generally, wide 95% LoA was found for the predicted IOL powers with the four IOL calculation formulas between the four devices.
The Anterion showed good repeatability of biometric measurements for most parameters. Good agreement among the four optical biometers was achieved for all the parameters except for CCT and the predicted IOL power. The AL values exhibited the best repeatability with Anterion and the best agreement among the biometers in our study.
研究 Anterion 的可重复性,并与其他三种光学生物测量仪比较眼生物测量和预测人工晶状体(IOL)度数的一致性。
记录 Anterion、IOLMaster 700、Lenstar LS 900 和 OA-2000 的平角膜曲率(Kf)、陡角膜曲率(Ks)、J0 和 45 向量、中央角膜厚度(CCT)、前房深度(ACD)和眼轴长度(AL)。采用 Hoffer Q、Holladay 1、SRK/T 和 Haigis 公式计算 IOL 度数。采用组内标准差(Sw)、重复性系数(RC)、变异系数(COV)和组内相关系数(ICC)评估重复性。采用 95%一致性界限评估四台生物测量仪之间的设备间一致性。
共纳入 101 名 101 只右眼。Anterion 对所有纳入的生物测量参数均具有良好的重复性,除 J45 外,所有参数的 COV≤0.30%,ICC≥0.930,J45 的重复性为中度(ICC 为 0.849)。四台设备的 Kf、Ks、J0、J45、ACD 和 AL 之间具有良好的一致性。通常,四种 IOL 计算公式在四台设备之间预测的 IOL 度数的 95%一致性界限较宽。
Anterion 对大多数参数的生物测量重复性较好。四台光学生物测量仪的所有参数之间均具有良好的一致性,除 CCT 和预测的 IOL 度数外。在我们的研究中,AL 值在 Anterion 中具有最佳的重复性,在生物测量仪中具有最佳的一致性。