Uster Stephanie, Topalli Zana, Sasse Thomas, Suter-Riniker Franziska, Barbani Maria Teresa
Institute for Infectious Diseases, University of Bern, Friedbühlstrasse 51, 3001 Bern, Switzerland.
J Clin Virol Plus. 2022 Aug;2(3):100095. doi: 10.1016/j.jcvp.2022.100095. Epub 2022 Jun 26.
SARS-CoV-2 antigen tests reliably detect individuals with high viral loads and provide an efficient diagnostic tool to manage the current SARS-CoV-2 pandemic However, mutations in SARS-CoV-2 variants of concerns that appeared after validation of most antigen tests might impact their diagnostic performance.
To assess the impact of the Omicron variant on the performance of the DiaSorin LIAISON SARS-CoV-2 antigen test, we evaluated its sensitivity and specificity on nasopharyngeal swabs (NPS) compared to rRT-PCR in the second and the Omicron pandemic wave in Switzerland.
A random selection of NPS from patients undergoing SARS-CoV-2 diagnostics by rRT-PCR were collected during the second and the Omicron pandemic wave and further analyzed by the LIAISON antigen test. Sensitivity and specificity compared to rRT-PCR were calculated.
Test performance did not change in the two investigated periods. The overall sensitivity of 75.8% in the second and 76.5% in the Omicron wave increased to 87.1% and 88.4%, excluding samples with rRT-PCR Ct-value >30. By lowering the cut-off from 200 TCID/ml to 62 TCID/ml to discriminate between negative and positive samples using a ROC-curve, the sensitivity resulted in 88.8% for the second and 93.3% for the Omicron pandemic wave. The specificity of the LIAISON antigen test was 100% in both collectives.
Omicron variant does not seem to affect the performance of the LIAISON antigen test. The WHO recommended sensitivity of ≥80% for antigen testing was fulfilled during both pandemic periods in samples with Ct-value <30 or by optimizing the assay cut-off.
严重急性呼吸综合征冠状病毒2(SARS-CoV-2)抗原检测能够可靠地检测出病毒载量高的个体,并为应对当前的SARS-CoV-2大流行提供了一种有效的诊断工具。然而,在大多数抗原检测验证之后出现的值得关注的SARS-CoV-2变异株中的突变可能会影响其诊断性能。
为评估奥密克戎变异株对DiaSorin LIAISON SARS-CoV-2抗原检测性能的影响,我们在瑞士第二波和奥密克戎大流行期间,比较了该检测对鼻咽拭子(NPS)与逆转录-聚合酶链反应(rRT-PCR)相比的敏感性和特异性。
在第二波和奥密克戎大流行期间,随机选取接受rRT-PCR进行SARS-CoV-2诊断的患者的NPS,并通过LIAISON抗原检测进一步分析。计算与rRT-PCR相比的敏感性和特异性。
在两个研究时期,检测性能没有变化。第二波时总体敏感性为75.8%,奥密克戎波时为76.5%,排除rRT-PCR Ct值>30的样本后,敏感性分别提高到87.1%和88.4%。通过使用ROC曲线将区分阴性和阳性样本的临界值从200 TCID/ml降低到62 TCID/ml,第二波时敏感性为88.8%,奥密克戎大流行波时为93.3%。LIAISON抗原检测在两个样本组中的特异性均为100%。
奥密克戎变异株似乎不影响LIAISON抗原检测的性能。在两个大流行时期,对于Ct值<30的样本或通过优化检测临界值,抗原检测达到了世界卫生组织建议的≥80%的敏感性。