Biosensors (Basel). 2022 Jun 30;12(7):479. doi: 10.3390/bios12070479.
Pulse wave velocity (PWV) measured at a specific artery location is called local PWV, which provides the elastic characteristics of arteries and indicates the degree of arterial stiffness. However, the large and cumbersome ultrasound probes require an appropriate sensor position and pressure maintenance, introducing usability constraints. In this paper, we developed a light (0.5 g) and thin (400 μm) flexible ultrasound array by encapsulating 1-3 composite piezoelectric transducers with a silicone elastomer. It can capture the distension waveforms of four arterial positions with a spacing of 10 mm and calculate the local PWV by multi-point fitting. This is illustrated by in vivo experiments, where the local PWV value of five normal subjects ranged from 3.07 to 4.82 m/s, in agreement with earlier studies. The beat-to-beat coefficient of variation (CV) is 12.0% ± 3.5%, showing high reliability. High reproducibility is shown by the results of two groups of independent measurements of three subjects (the error between the mean values is less than 0.3 m/s). These properties of the developed flexible ultrasound array enable the bandage-like application of local PWV monitoring to skin surfaces.
脉搏波速度(PWV)在特定动脉位置测量称为局部 PWV,它提供了动脉的弹性特征,并表明了动脉僵硬的程度。然而,大型且笨重的超声波探头需要合适的传感器位置和压力保持,这引入了可用性限制。在本文中,我们通过用硅橡胶弹性体封装 1-3 个复合压电换能器,开发了一种重量轻(0.5 克)、厚度薄(400μm)的柔性超声阵列。它可以捕获四个动脉位置的膨胀波形,间距为 10mm,并通过多点拟合计算局部 PWV。通过体内实验说明了这一点,其中五个正常受试者的局部 PWV 值范围为 3.07 至 4.82m/s,与早期研究一致。心跳到心跳的变异系数(CV)为 12.0%±3.5%,显示出很高的可靠性。三个受试者的两组独立测量结果(平均值之间的误差小于 0.3m/s)显示出很高的重现性。这种开发的柔性超声阵列的特性使得局部 PWV 监测能够像绷带一样应用于皮肤表面。