在健康的孕晚期孕妇队列中,比较旋转血栓弹力图(ROTEM)Delta设备与基于试剂盒的血栓弹力图6s及Quantra的性能。

Comparison between the Rotational Thromboelastometry (ROTEM) Delta device against the Cartridge-based Thromboelastography 6s and Quantra in a healthy third trimester pregnant cohort.

作者信息

Gonzalez-Fiol Antonio, Fardelmann Kristen L, Yanez David, Salimi Nayema, Mancini Peter, Alian Aymen

机构信息

Department of Anesthesiology, Yale School of Medicine, 333 Cedar Street, TMP 3, 06510, New Haven, CT, USA.

出版信息

J Clin Monit Comput. 2023 Feb;37(1):267-273. doi: 10.1007/s10877-022-00888-0. Epub 2022 Jul 29.

Abstract

Rotational Thromboelastometry (ROTEM) Delta has been described in several postpartum hemorrhage algorithms, but this device requires pipetting and careful mixing of reagents to initiate the clotting reaction. In contrast, thromboelastography (TEG 6s) and the Quantra devices operate utilizing an automated pre-mixed cartridge that only requires a blood sample to start the clot strength analysis. We compared the correlation between 3 point of care viscoelastic testing (POCVT) devices to laboratory Clauss fibrinogen and platelets, their inter-device correlation, and the total running time difference between Quantra and ROTEM. A high correlation was noted between the Clauss fibrinogen and the fibrinogen parameters from ROTEM (r = 0.76-0.84, P < 0.0001), TEG6s (r = 0.71, P < 0.0001) and Quantra (r = 0.72, P = 0.0001). A moderate correlation between laboratory platelets and the ROTEM (r = 0.54;0.45, P < 0.0001; P = 0.0013) and Quantra (r = 0.66, P = 0.0001) parameters was noted. The inter-device correlation showed to be high when comparing the fibrinogen parameters of TEG6s and Quantra to that of ROTEM (r = 0.88 and 0.74, P < 0.0001, respectively). In contrast, a moderate correlation was noted between the platelet parameters of Quantra and ROTEM (r = 0.51, p = 0.0036). The Quantra device resulted 20.9 min (95% CI -0.2 to 4.7, P = 0.07) faster than the ROTEM if the warming and pipetting of reagents of the latter were considered. All the POCVT devices demonstrated a high correlation to laboratory Clauss fibrinogen, making each beneficial for the early recognition and management of hypofibrinogenemia.

摘要

旋转血栓弹力图(ROTEM)Delta已在多种产后出血算法中有所描述,但该设备需要移液并仔细混合试剂以启动凝血反应。相比之下,血栓弹力描记法(TEG 6s)和Quantra设备利用自动预混试剂盒进行操作,该试剂盒仅需一份血样即可开始凝血强度分析。我们比较了三种床旁粘弹性检测(POCVT)设备与实验室Clauss纤维蛋白原和血小板之间的相关性、设备间的相关性以及Quantra和ROTEM之间的总运行时间差异。结果发现,Clauss纤维蛋白原与ROTEM(r = 0.76 - 0.84,P < 0.0001)、TEG6s(r = 0.71,P < 0.0001)和Quantra(r = 0.72,P = 0.0001)的纤维蛋白原参数之间存在高度相关性。实验室血小板与ROTEM(r = 0.54;0.45,P < 0.0001;P = 0.0013)和Quantra(r = 0.66,P = 0.0001)参数之间存在中度相关性。将TEG6s和Quantra的纤维蛋白原参数与ROTEM的进行比较时,设备间相关性较高(r分别为0.88和0.7,4,P < 0.0001)。相比之下,Quantra和ROTEM的血小板参数之间存在中度相关性(r = 0.51,p = 0.0036)。如果考虑ROTEM试剂的预热和移液操作,Quantra设备比ROTEM快20.9分钟(95% CI -0.2至4.7,P = 0.07)。所有POCVT设备与实验室Clauss纤维蛋白原均显示出高度相关性,这使得它们在纤维蛋白原血症的早期识别和管理中都具有益处。

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验