Gonzalez-Fiol Antonio, Fardelmann Kristen L, Yanez David, Salimi Nayema, Mancini Peter, Alian Aymen
Department of Anesthesiology, Yale School of Medicine, 333 Cedar Street, TMP 3, 06510, New Haven, CT, USA.
J Clin Monit Comput. 2023 Feb;37(1):267-273. doi: 10.1007/s10877-022-00888-0. Epub 2022 Jul 29.
Rotational Thromboelastometry (ROTEM) Delta has been described in several postpartum hemorrhage algorithms, but this device requires pipetting and careful mixing of reagents to initiate the clotting reaction. In contrast, thromboelastography (TEG 6s) and the Quantra devices operate utilizing an automated pre-mixed cartridge that only requires a blood sample to start the clot strength analysis. We compared the correlation between 3 point of care viscoelastic testing (POCVT) devices to laboratory Clauss fibrinogen and platelets, their inter-device correlation, and the total running time difference between Quantra and ROTEM. A high correlation was noted between the Clauss fibrinogen and the fibrinogen parameters from ROTEM (r = 0.76-0.84, P < 0.0001), TEG6s (r = 0.71, P < 0.0001) and Quantra (r = 0.72, P = 0.0001). A moderate correlation between laboratory platelets and the ROTEM (r = 0.54;0.45, P < 0.0001; P = 0.0013) and Quantra (r = 0.66, P = 0.0001) parameters was noted. The inter-device correlation showed to be high when comparing the fibrinogen parameters of TEG6s and Quantra to that of ROTEM (r = 0.88 and 0.74, P < 0.0001, respectively). In contrast, a moderate correlation was noted between the platelet parameters of Quantra and ROTEM (r = 0.51, p = 0.0036). The Quantra device resulted 20.9 min (95% CI -0.2 to 4.7, P = 0.07) faster than the ROTEM if the warming and pipetting of reagents of the latter were considered. All the POCVT devices demonstrated a high correlation to laboratory Clauss fibrinogen, making each beneficial for the early recognition and management of hypofibrinogenemia.
旋转血栓弹力图(ROTEM)Delta已在多种产后出血算法中有所描述,但该设备需要移液并仔细混合试剂以启动凝血反应。相比之下,血栓弹力描记法(TEG 6s)和Quantra设备利用自动预混试剂盒进行操作,该试剂盒仅需一份血样即可开始凝血强度分析。我们比较了三种床旁粘弹性检测(POCVT)设备与实验室Clauss纤维蛋白原和血小板之间的相关性、设备间的相关性以及Quantra和ROTEM之间的总运行时间差异。结果发现,Clauss纤维蛋白原与ROTEM(r = 0.76 - 0.84,P < 0.0001)、TEG6s(r = 0.71,P < 0.0001)和Quantra(r = 0.72,P = 0.0001)的纤维蛋白原参数之间存在高度相关性。实验室血小板与ROTEM(r = 0.54;0.45,P < 0.0001;P = 0.0013)和Quantra(r = 0.66,P = 0.0001)参数之间存在中度相关性。将TEG6s和Quantra的纤维蛋白原参数与ROTEM的进行比较时,设备间相关性较高(r分别为0.88和0.7,4,P < 0.0001)。相比之下,Quantra和ROTEM的血小板参数之间存在中度相关性(r = 0.51,p = 0.0036)。如果考虑ROTEM试剂的预热和移液操作,Quantra设备比ROTEM快20.9分钟(95% CI -0.2至4.7,P = 0.07)。所有POCVT设备与实验室Clauss纤维蛋白原均显示出高度相关性,这使得它们在纤维蛋白原血症的早期识别和管理中都具有益处。