Magnetically controlled capsule endoscopy (MCCE) has recently increasingly been used for gastric examination. However, the image quality and esophageal observation is suboptimal. We developed a novel wired transmission magnetically controlled capsule endoscopy (WT-MCCE) system and evaluated its feasibility through in vitro and in vivo experiments.
A plastic stomach model and a pathological upper gastrointestinal model were used to evaluate the performance of WT-MCCE in vitro experiments. Twice of examination in the two in vitro models by WT-MCCE were performed by 5 endoscopists who were experienced in performing wireless capsule endoscopy. The examination of traditional gastroscopy (Olympus, GIF-HQ290) in the pathological upper gastrointestinal model was set as the control. In vivo experiments were performed in a live canine model by 3 endoscopists, in which WT-MCCE was inserted with the assistance of gastroscopy. Measurements included maneuverability, examination time, visualization of gastric mucosa, image quality and diagnostic accuracy.
WT-MCCE showed good performance in both in vitro and in vivo experiments with excellent visualization of mucosa (75-100%). The mean operation time is 17.6 ± 2.7 min, 22.3 ± 1.9 min and 29.3 ± 3.4 min in three models, respectively. In pathological upper gastrointestinal model, all lesions, including esophageal varices, one polyp, one foreign body, two gastric ulcers and one duodenal ulcer, were detected by both WT-MCCE and traditional gastroscopy by all endoscopists. For the observation of esophagus and stomach in the canine model, WT-MCCE also showed excellent maneuverability and good image quality.
The novel WT-MCCE system performed well in evaluating upper gastrointestinal landmarks and lesions in two in vitro models, and showed good performance in a canine model. WT-MCCE may be potentially useful for diagnosis of esophageal and gastric diseases.
磁控胶囊内镜(MCCE)最近越来越多地用于胃部检查,但图像质量和食管观察效果不理想。我们开发了一种新型的有线传输磁控胶囊内镜(WT-MCCE)系统,并通过体外和体内实验评估了其可行性。
使用塑料胃模型和病理性上消化道模型进行体外实验,由 5 名具有无线胶囊内镜检查经验的内镜医生对 WT-MCCE 在两个体外模型中的检查结果进行了两次检查。病理性上消化道模型中传统胃镜(奥林巴斯,GIF-HQ290)检查作为对照。由 3 名内镜医生在活体犬模型中进行体内实验,在胃镜的协助下插入 WT-MCCE。测量指标包括操控性、检查时间、胃黏膜可视化、图像质量和诊断准确性。
WT-MCCE 在体外和体内实验中均表现出良好的性能,胃黏膜可视化效果极佳(75%-100%)。在三种模型中,平均操作时间分别为 17.6±2.7min、22.3±1.9min 和 29.3±3.4min。在病理性上消化道模型中,所有病变,包括食管静脉曲张、1 个息肉、1 个异物、2 个胃溃疡和 1 个十二指肠溃疡,均被 WT-MCCE 和传统胃镜由所有内镜医生检出。对于犬模型中食管和胃的观察,WT-MCCE 也表现出良好的操控性和良好的图像质量。
新型 WT-MCCE 系统在两种体外模型中对上消化道标志物和病变的评估表现良好,在犬模型中也表现出良好的性能。WT-MCCE 可能对食管和胃部疾病的诊断具有潜在的应用价值。