使用四种设备进行白对白测量及其对眼内晶状体(ICL)尺寸测定的比较。
Comparison of white-to-white measurements using four devices and their determination of ICL sizing.
作者信息
Ang Robert Edward T, Reyes Edward Kenneth F, Ayuyao Fernando Amado J, Umali Maria Isabel N, Cruz Emerson M
机构信息
Asian Eye Institute, Phinma Plaza, Rockwell Center, 8Th Floor, 1200, Makati, Philippines.
Cardinal Santos Medical Center, 10 Wilson St., Greenhills, 1502, San Juan, Philippines.
出版信息
Eye Vis (Lond). 2022 Oct 2;9(1):36. doi: 10.1186/s40662-022-00308-z.
BACKGROUND
To compare the measurements obtained from the Orbscan II, IOLMaster 700, Pentacam AXL, and Castroviejo caliper and their effects on calculating the recommended implantable collamer lens (ICL) size and postoperative vault measurements.
METHODS
This is a retrospective cross-sectional study of patients who underwent ICL surgery by a single surgeon from March 1, 2018 to July 31, 2021. Records were reviewed for the anterior chamber depth (ACD) and white-to-white (WTW) measurements obtained from the Orbscan II, IOLMaster 700, Pentacam AXL, and Castroviejo caliper (WTW only). These were used to calculate the recommended ICL size. The actual ICL size implanted, and vault measurements obtained one month postoperatively were also collected.
RESULTS
One hundred seven eyes with a mean age of 27.9 ± 7.7 years were included in the study. Mean WTW measurements were significantly different between devices (P < 0.0001), with the IOLMaster 700 having the highest value (12.14 ± 0.04 mm) and the caliper having the lowest value (11.45 ± 0.04 mm). Mean ACD measurements were the lowest in Orbscan II (3.12 ± 0.25 mm) and the highest in Pentacam AXL (3.16 ± 0.24 mm). The Pentacam AXL produced an ICL size similar to the Orbscan in 69.2% of eyes. The IOLMaster yielded an ICL measurement one size larger than Orbscan-based calculations in 64.5% of eyes. Using the Orbscan WTW and ACD, the desired vault of 0.25 to 0.75 mm and 0.25 to 1.00 mm was achieved in 70% and 91% of eyes, respectively. Substituting caliper WTW to IOLMaster 700 or Pentacam AXL WTW increases the percentage of achieving the desired vault to 80%, similar to the Orbscan.
CONCLUSIONS
The Orbscan II, IOLMaster 700, and Pentacam AXL cannot be used interchangeably for calculating ICL sizing. Combining the WTW from caliper measurement with the ACD of the IOLMaster 700 or Pentacam AXL could improve ICL sizing and achieve a higher percentage of eyes with the desired vault.
背景
比较Orbscan II、IOLMaster 700、Pentacam AXL和Castroviejo卡尺所获得的测量值,以及它们对计算推荐的可植入式角膜接触镜(ICL)尺寸和术后房拱高度测量值的影响。
方法
这是一项回顾性横断面研究,研究对象为2018年3月1日至2021年7月31日由同一位外科医生进行ICL手术的患者。回顾记录中从Orbscan II、IOLMaster 700、Pentacam AXL和Castroviejo卡尺(仅用于测量白到白距离)获得的前房深度(ACD)和白到白距离(WTW)测量值。这些测量值用于计算推荐的ICL尺寸。还收集了实际植入的ICL尺寸以及术后1个月获得的房拱高度测量值。
结果
该研究纳入了107只眼睛,平均年龄为27.9±7.7岁。不同设备之间的平均WTW测量值存在显著差异(P<0.0001),其中IOLMaster 700的值最高(12.14±0.04mm),卡尺的值最低(11.45±0.04mm)。平均ACD测量值在Orbscan II中最低(3.12±0.25mm),在Pentacam AXL中最高(3.16±0.24mm)。Pentacam AXL在69.2%的眼睛中得出的ICL尺寸与Orbscan相似。在64.5%的眼睛中,IOLMaster得出的ICL测量值比基于Orbscan的计算结果大一个尺寸。使用Orbscan的WTW和ACD,分别在70%和91%的眼睛中实现了0.25至0.75mm和0.25至1.00mm的理想房拱高度。将卡尺测量的WTW替换为IOLMaster 700或Pentacam AXL测量的WTW,可使达到理想房拱高度的眼睛比例提高到与Orbscan相似的80%。
结论
Orbscan II、IOLMaster 700和Pentacam AXL在计算ICL尺寸时不能互换使用。将卡尺测量的WTW与IOLMaster 700或Pentacam AXL的ACD相结合,可以改善ICL尺寸的计算,并使更多眼睛达到理想的房拱高度。
相似文献
Eye Vis (Lond). 2022-10-2
J Cataract Refract Surg. 2022-11-1
Clin Exp Optom. 2015-5
BMJ Open Ophthalmol. 2022-3-8
引用本文的文献
BMC Ophthalmol. 2025-8-27
Int J Ophthalmol. 2025-5-18
Graefes Arch Clin Exp Ophthalmol. 2025-4-1
Med Hypothesis Discov Innov Ophthalmol. 2024-8-14
Indian J Ophthalmol. 2024-7-1
Medicine (Baltimore). 2024-5-17
本文引用的文献
PLoS One. 2020-11-4
Indian J Ophthalmol. 2019-3
Am J Ophthalmol. 2017-12-30
Middle East Afr J Ophthalmol. 2016
J Cataract Refract Surg. 2015-3
Expert Rev Med Devices. 2015-1