由于瞳孔间距测量的偏差,后房有晶状体眼人工晶状体选择中的混杂尺寸问题。

Confounding sizing in posterior chamber phakic lens selection due to white-to-white measurement bias.

机构信息

Department of Ophthalmology (Qvision), Vithas Virgen del Mar Hospital, 04120; Department of Ophthalmology, Torrecárdenas Hospital Complex, 04009, Almería, Spain.

Department of Ophthalmology (Qvision), Vithas Virgen del Mar Hospital, 04120, Almería, Spain.

出版信息

Indian J Ophthalmol. 2019 Mar;67(3):344-349. doi: 10.4103/ijo.IJO_613_18.

Abstract

PURPOSE

To assess the agreement in the white-to-white (WTW) measurement with two different devices, the reproducibility and the probability of confusing sizing (PCS) in selecting a different implantable collamer lens (ICL).

STUDY DESIGN

Retrospective observational case series.

METHODS

Images of 192 eyes were captured with both devices. The WTW was measured automatically (OA) and manually (OM) with the Orbscan and Keratograph (KA and KM) by one examiner who repeated a total of four measures. A second examiner conducted a single manual measure for each device over the same image. The ICL sizing was computed for each measure of WTW and the PCS was calculated as the percentage of cases for which the confronted or repeated measure resulted in a different size of the ICL. The critical WTWs with highest PCS were identified.

RESULTS

KM overestimated the WTW versus OM in 0.13 ± 0.18 mm (P < 0.001) but not in the automated method comparison, 0.01 ± 0.19 mm (P = 0.58). Inter-examiner reproducibility (R) was higher with OM than with KM, and the intra-examiner R decreased with the average of two measures in both cases. The PCS was higher with the increase of mean differences, the limits of agreement (LoAs), and R. WTWs from 11.1 to 11.2 mm, 11.6 to 11.7 mm, and 12.3 to 12.4 mm resulted in higher PCS.

CONCLUSION

The mean difference is not enough to apply conversions between devices and the LoAs and R should be considered. Special attention should be taken for WTWs with higher PCS.

摘要

目的

评估两种不同设备的白到白(WTW)测量值之间的一致性、重复性和混淆选择不同可植入式隐形眼镜(ICL)尺寸的概率(PCS)。

研究设计

回顾性观察性病例系列。

方法

使用两种设备拍摄 192 只眼睛的图像。WTW 由一名检查者自动(OA)和手动(OM)进行测量,使用 Orbscan 和角膜地形图(KA 和 KM),共重复测量四次。第二名检查者对同一图像的每种设备进行一次手动测量。为每个 WTW 测量值计算 ICL 尺寸,并将 PCS 计算为面对或重复测量导致 ICL 尺寸不同的病例百分比。确定具有最高 PCS 的临界 WTW。

结果

与 OM 相比,KM 高估了 WTW,差值为 0.13 ± 0.18mm(P < 0.001),但在自动方法比较中,差值为 0.01 ± 0.19mm(P = 0.58)。OM 的检查者间重复性(R)高于 KM,两种情况下,平均两次测量的 R 都随 R 降低。PCS 随平均差值、一致性界限(LoA)和 R 的增加而增加。11.1 至 11.2mm、11.6 至 11.7mm 和 12.3 至 12.4mm 的 WTW 导致更高的 PCS。

结论

平均值差异不足以在设备之间进行转换,应考虑 LoA 和 R。对于 PCS 较高的 WTW,应特别注意。