Porter Paul, Zhou Huaqiong, Schneider Brooke, Choveaux Jennifer, Bear Natasha, Della Phillip, Jones Kym
Department of Paediatrics, Joondalup Health Campus, Perth, WA, Australia.
Faculty of Health Science, Curtin University, Perth, WA, Australia.
NPJ Digit Med. 2022 Nov 3;5(1):167. doi: 10.1038/s41746-022-00714-6.
Fetal Cardiography is usually performed using in-hospital Cardiotocographic (CTG) devices to assess fetal wellbeing. New technologies may permit home-based, self-administered examinations. We compared the accuracy, clinical interpretability, and user experience of a patient-administered, wireless, fetal heartbeat monitor (HBM) designed for home use, to CTG. Initially, participants had paired HBM and CTG examinations performed in the clinic. Women then used the HBM unsupervised and rated the experience. Sixty-three women had paired clinic-based HBM and CTG recordings, providing 6982 fetal heart rate measures for point-to-point comparison from 126 min of continuous recording. The accuracy of the HBM was excellent, with limits of agreement (95%) for mean fetal heart rate (FHR) between 0.72 and -1.78 beats per minute. The FHR was detected on all occasions and confirmed to be different from the maternal heart rate. Both methods were equally interpretable by Obstetricians, and had similar signal loss ratios. Thirty-four (100%) women successfully detected the FHR and obtained clinically useful cardiographic data using the device at home unsupervised. They achieved the required length of recording required for non-stress test analysis. The monitor ranked in the 96-100 percentile for usability and learnability. The HBM is as accurate as gold-standard CTG, and provides equivalent clinical information enabling use in non-stress test analyses conducted outside of hospitals. It is usable by expectant mothers with minimal training.
胎儿心动描记术通常使用医院内的胎心监护仪(CTG)来评估胎儿健康状况。新技术可能允许在家中进行自我管理的检查。我们将一种为家庭使用设计的、患者可自行操作的无线胎儿心跳监测仪(HBM)与CTG在准确性、临床可解释性和用户体验方面进行了比较。最初,参与者在诊所进行了HBM和CTG的配对检查。然后,女性在无人监督的情况下使用HBM,并对体验进行评分。63名女性有基于诊所的HBM和CTG配对记录,从126分钟的连续记录中提供了6982次胎儿心率测量值用于点对点比较。HBM的准确性极佳,平均胎儿心率(FHR)的一致性界限(95%)在每分钟0.72次至 -1.78次之间。所有情况下都检测到了FHR,并确认其与母亲心率不同。两种方法对产科医生来说同样易于解释,且信号丢失率相似。34名(100%)女性在无人监督的情况下在家中使用该设备成功检测到了FHR,并获得了临床上有用的心动描记数据。她们达到了无应激试验分析所需的记录时长。该监测仪在可用性和易学性方面排名在第96 - 100百分位。HBM与金标准CTG一样准确,并提供等效的临床信息,使其可用于医院外进行的无应激试验分析。它对于准妈妈来说,经过最少的培训即可使用。