[Comparison of Doppler echocardiographic methods of determining cardiac minute volume].

作者信息

Dittmann H, Voelker W, Karsch K R, Seipel L

机构信息

Medizinische Klinik III, Eberhard-Karls-Universität Tübingen.

出版信息

Z Kardiol. 1987 Jul;76(7):433-8.

PMID:3673165
Abstract

In 40 patients without valvular disease, cardiac output was determined by pulsed Doppler echocardiography and thermodilution simultaneously. The sample volume was located in the center of the mitral valve ring, at the tips of the mitral leaflets and in the left ventricular outflow tract, directly proximal to the aortic valve leaflets. Circular cross-sectional areas of the mitral valve ring, aortic ring and bulbus of the aorta were calculated from the M-mode and two dimensional echocardiographic diameters. The mitral orifice was assumed to be an ellipse with varying short axes, determined as the mean diastolic leaflet separation in the M-mode and a constant long axis, derived from the maximal mitral orifice area or mitral ring diameter. Cardiac output was calculated by multiplying time-velocity integrals with different areas and heart rate. Cardiac output, measured by the thermodilution technique, ranged from 4.0 l/min to 10.2 l/min. Cardiac output determined by the different Doppler methods correlated significantly with the thermodilution measurements. Cardiac output measurements in the left ventricular outflow tract provided the best correlation coefficient (0.93) and a standard error of the estimate of 0.589 l/min, when the circular flow area was derived from the M-mode echo of the aortic ring.(ABSTRACT TRUNCATED AT 250 WORDS)

摘要

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验