[Noninvasive measurement of cardiac output by pulsed Doppler echocardiography. Correlation with thermodilution].

作者信息

Dericbourg C, Tribouilloy C, Kugener H, Avinee P, Rey J L, Lesbre J P

机构信息

CHU, Amiens.

出版信息

Arch Mal Coeur Vaiss. 1990 Feb;83(2):237-44.

PMID:2106860
Abstract

Cardiac output was measured simultaneously by pulsed Doppler echocardiography and thermodilution in 22 patients, 18 of whom also underwent atrial pacing at different rates to give a total of 42 different measurements. The aortic diameter was measured firstly at the aortic ring at the level of insertion of the aortic cusps and then at the point of maximum separation of the valve cups in the left parasternal long-axis view. The aortic velocities were recorded in the apical 5-chamber view immediately below the level of the aortic valve. The correlations obtained at the aortic ring (R1) and at the point of maximum separation of the valve cusps (R2) were 0.77 (y = 0.67x + 1.17: standard error = 0.81 l/m) and 0.64 (y = 0.56x + 0.87; standard error = 1.01 l/mn) respectively. The correlations were much better when 7 technically unsatisfactory measurements were excluded (R2 = 0.76: y = 0.59x + 0.74: standard error = 0.79 l/mn) (R1 = 0.87: y = 0.72x + 1.04: standard error = 0.65 l/mn). THe correlations of stroke volume measured at aortic ring level also improved from r = 0.82 (y = 0.75x + 7.29: standard error = 8.9 ml) to r = 0.89 (y = 0.78x + 7.38: standard error = 7.3 ml). The measurement of cardiac output by pulsed Doppler echocardiography in the aortic root seems to be reliable. The correlations of the values of stroke volume and cardiac output with the thermodilution method are good, allowing detection of beat-to-beat variations of cardiac output, in suitable patients in the hands of experienced operators.

摘要

相似文献

1
[Noninvasive measurement of cardiac output by pulsed Doppler echocardiography. Correlation with thermodilution].
Arch Mal Coeur Vaiss. 1990 Feb;83(2):237-44.
2
[Noninvasive measurement of cardiac output using two-dimensional Doppler echocardiography and analysis of sources of error].[使用二维多普勒超声心动图无创测量心输出量及误差来源分析]
J Cardiol. 1987 Mar;17(1):139-48.
3
[Comparison of Doppler echocardiographic methods of determining cardiac minute volume].
Z Kardiol. 1987 Jul;76(7):433-8.
4
[Transesophageal Doppler echocardiographic measurement of cardiac output using mitral anulus method].[经食管多普勒超声心动图采用二尖瓣环法测量心输出量]
Kokyu To Junkan. 1991 Jul;39(7):665-8.
5
[Doppler echocardiography measurement of the regurgitation fraction in patients with aortic valve insufficiency].[多普勒超声心动图测量主动脉瓣关闭不全患者的反流分数]
Z Kardiol. 1989 Dec;78(12):771-6.
6
Reliability of Doppler echocardiographic stroke volume measurement.
Bull Tokyo Med Dent Univ. 1992 Sep;39(3):55-62.
7
[Noninvasive determination of cardiac output by continuous wave Doppler technique using apical approach].[采用心尖入路的连续波多普勒技术无创测定心输出量]
Hokkaido Igaku Zasshi. 1991 Mar;66(2):170-8.
8
Cardiac output measurement during infrarenal aortic surgery: echo-esophageal Doppler versus thermodilution catheter.肾下腹主动脉手术期间的心输出量测量:经食管超声多普勒与热稀释导管法的比较
J Cardiothorac Vasc Anesth. 2006 Feb;20(1):26-30. doi: 10.1053/j.jvca.2005.07.029.
9
[Doppler echocardiography determination of the minute heart volume in ergometric stress].[运动负荷时通过多普勒超声心动图测定每分心输出量]
Z Kardiol. 1987 Jun;76(6):346-50.
10
Transoesophageal echocardiography accurately detects cardiac output variation: a prospective comparison with thermodilution in cardiac surgery.经食管超声心动图可准确检测心输出量变化:与心脏手术中热稀释法的前瞻性比较
Eur J Anaesthesiol. 2008 Feb;25(2):135-43. doi: 10.1017/S0265021507001354. Epub 2007 Aug 2.

引用本文的文献

1
Carotid Doppler Imaging as a Marker for Fluid Responsiveness.颈动脉多普勒成像作为液体反应性的标志物
J Clin Med. 2025 Sep 22;14(18):6657. doi: 10.3390/jcm14186657.
2
Comparison of cardiac index measurements in intensive care patients using continuous wave vs. pulsed wave echo-Doppler compared to pulse contour cardiac output.在重症监护患者中,使用连续波与脉冲波回声多普勒测量心脏指数,并与脉搏轮廓心输出量进行比较。
Intensive Care Med Exp. 2023 Apr 28;11(1):23. doi: 10.1186/s40635-023-00499-2.
3
From left ventricular ejection fraction to cardiac hemodynamics: role of echocardiography in evaluating patients with heart failure.
从左心室射血分数到心脏血流动力学:超声心动图在心力衰竭患者评估中的作用。
Heart Fail Rev. 2020 Mar;25(2):217-230. doi: 10.1007/s10741-019-09826-w.
4
Transthoracic echocardiography: an accurate and precise method for estimating cardiac output in the critically ill patient.经胸超声心动图:一种用于评估危重症患者心输出量的准确、精确的方法。
Crit Care. 2017 Jun 9;21(1):136. doi: 10.1186/s13054-017-1737-7.
5
Can dynamic indicators help the prediction of fluid responsiveness in spontaneously breathing critically ill patients?动态指标能否有助于预测自主呼吸的危重症患者的液体反应性?
Intensive Care Med. 2007 Jul;33(7):1117-1124. doi: 10.1007/s00134-007-0644-9. Epub 2007 May 17.
6
Non-invasive measurement of the regurgitant fraction by pulsed Doppler echocardiography in isolated pure mitral regurgitation.经脉冲多普勒超声心动图对单纯二尖瓣反流患者反流分数的无创测量。
Br Heart J. 1991 Oct;66(4):290-4. doi: 10.1136/hrt.66.4.290.