Calibration of a novel rebound tonometer (Tono-Vera® vet) in the porcine eye.

机构信息

Department of Veterinary Clinical Sciences, School of Veterinary Medicine, Louisiana State University, Baton Rouge, Louisiana, USA.

出版信息

Vet Ophthalmol. 2023 May;26(3):268-272. doi: 10.1111/vop.13067. Epub 2023 Feb 7.

Abstract

OBJECTIVE

Porcine models of ocular disease are becoming increasingly utilized. A recently commercialized ocular tonometer, the Reichert Tono-Vera® Vet, has not been evaluated for use in pigs. The purpose of this study was to calibrate this device for use in porcine eyes and to determine which settings are most appropriate for use in pigs.

PROCEDURE

The anterior chambers of five freshly enucleated normal porcine eyes were cannulated then connected to a reservoir of balanced salt solution and a physiologic monitor. Triplicate measurements were obtained with the four available settings: dog, cat, horse, and rabbit at intraocular pressures ranging from 5- to 80 mmHg. Bland-Altman analysis was utilized to determine bias and 95% limits of agreement for each setting.

RESULTS

There was a strong positive linear regression trend for all settings (dog r  = 0.986, horse r  = 0.947, cat r  = 0.977, and rabbit r  = 0.982). The linear regression equations for the dog, horse, cat, and rabbit setting were y = 1.0168x - 2.6128, y = 0.8743x - 3.4959, y = 0.9394x - 7.3188, and y = 1.1082x - 3.4077. The average bias and 95% limits of agreement for dog, horse, cat, and rabbit settings were - 2.00, -8.32, -9.58, and 0.57 mmHg, and (-7.52, 3.53), (-19.00, 2.37), (-16.66, -2.50), and (-7.79, 8.93), in mmHg.

CONCLUSION

The Tono-Vera® Vet dog setting was most accurate and precise setting compared to true intraocular pressures. This setting is likely to be appropriate for in vivo use in pigs, with the appropriate correction formula applied.

摘要

目的

眼部疾病的猪模型越来越多地被应用。最近商业化的眼部眼压计 Reichert Tono-Vera® Vet 尚未在猪身上进行评估。本研究的目的是对该设备进行校准,使其适用于猪眼,并确定最适合猪使用的设置。

过程

五只新鲜取出的正常猪眼的前房被插管,然后连接到平衡盐溶液和生理监测器的储液器。在眼压范围为 5 至 80mmHg 时,使用四种可用设置(狗、猫、马和兔)进行了三次重复测量。Bland-Altman 分析用于确定每个设置的偏倚和 95%一致性界限。

结果

所有设置均呈现出强烈的正线性回归趋势(狗 r  = 0.986,马 r  = 0.947,猫 r  = 0.977,兔 r  = 0.982)。狗、马、猫和兔设置的线性回归方程分别为 y  = 1.0168x - 2.6128,y  = 0.8743x - 3.4959,y  = 0.9394x - 7.3188,y  = 1.1082x - 3.4077。狗、马、猫和兔设置的平均偏差和 95%一致性界限分别为-2.00、-8.32、-9.58 和 0.57mmHg,以及(-7.52,3.53)、(-19.00,2.37)、(-16.66,-2.50)和(-7.79,8.93)mmHg。

结论

与真实眼压相比,Tono-Vera® Vet 狗设置是最准确和最精确的设置。在应用适当的校正公式后,该设置可能适用于猪的体内使用。

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验