新型回弹眼压计(Tono-Vera® Vet)在正常离体兔眼的验证。

Validation of a novel rebound tonometer (Tono-Vera® Vet) in normal ex vivo rabbit eyes.

机构信息

Department of Veterinary Clinical Sciences, School of Veterinary Medicine, Louisiana State University, Baton Rouge, Louisiana, USA.

出版信息

Vet Ophthalmol. 2023 Nov;26(6):555-559. doi: 10.1111/vop.13142. Epub 2023 Sep 1.

Abstract

OBJECTIVE

The purpose of this study was to validate the use of the Reichert Tono-Vera® Vet tonometer rabbit setting in normal ex vivo rabbit eyes and to compare the rabbit setting to the dog, cat, and horse settings of this tonometer.

PROCEDURE

Six freshly enucleated normal rabbit eyes were cannulated and connected to a fluid reservoir and physiologic monitor. Triplicate measurements were obtained with the four available settings: dog, cat, horse, and rabbit at various intraocular pressures (IOP) ranging from 5 to 80 mmHg. Bland-Altman analysis was utilized to determine bias and 95% limits of agreement for each setting.

RESULTS

Linear regression equations for the dog, horse, cat, and rabbit settings were y = 0.8101x + 2.5058, y = 0.7594x - 3.4673, y = 0.6635x + 0.3021, and y = 0.8935x + 1.3295, respectively. All settings demonstrated strong positive linear trends (dog r  = 0.9644, horse r  = 0.9456, cat r  = 0.9309, and rabbit r  = 0.9558). Bland-Altman plots revealed that the average bias and 95% limits of agreement (mmHg) were -4.73, -12.65, -12.86 and -2.73 and (-15.31, 5.86), (-29.03, 3.74), (-25.67, -0.05), and (-12.21, 6.76) for the dog, horse, cat, and rabbit settings, respectively.

CONCLUSION

The Tono-Vera® Vet rabbit setting provided the most accurate and precise measurements compared with the other settings, but slightly underestimated actual IOP, especially as IOP was increased. This tonometer, using the rabbit setting, is likely to be appropriate for the estimation of IOP in rabbits with the appropriate correction formula applied.

摘要

目的

本研究旨在验证 Reichert Tono-Vera® Vet 兔眼压计在正常离体兔眼的使用,并比较该眼压计的兔眼压计设置与犬、猫和马眼压计设置。

方法

将 6 只新鲜取出的正常兔眼进行套管,并连接到液体储液器和生理监测器上。在 5 至 80mmHg 的不同眼内压 (IOP) 范围内,使用四种可用设置(犬、猫、马和兔)进行了三次重复测量。采用 Bland-Altman 分析确定每种设置的偏差和 95%一致性界限。

结果

犬、马、猫和兔眼压计设置的线性回归方程分别为 y=0.8101x+2.5058、y=0.7594x-3.4673、y=0.6635x+0.3021 和 y=0.8935x+1.3295。所有设置均表现出强正线性趋势(犬 r=0.9644、马 r=0.9456、猫 r=0.9309 和兔 r=0.9558)。Bland-Altman 图显示,平均偏差和 95%一致性界限(mmHg)分别为-4.73、-12.65、-12.86 和-2.73 和(-15.31,5.86)、(-29.03,3.74)、(-25.67,-0.05)和(-12.21,6.76),用于犬、马、猫和兔眼压计设置。

结论

与其他设置相比,Tono-Vera® Vet 兔眼压计设置提供了最准确和最精确的测量值,但略微低估了实际的 IOP,尤其是在 IOP 升高时。使用兔眼压计设置,该眼压计可能适用于适当校正公式应用的兔眼 IOP 的估计。