Herrero-Climent Mariano, Falcao Artur, Tondela Joao, Brizuela Aritza, Rios-Carrasco Blanca, Gil Javier
Porto Dental Institute, Av. de Montevideu 810, 4150-518 Porto, Portugal.
Centre for Innovation and Research in Oral Sciences (CIROS), Faculty of Medicine, University of Coimbra, Rua Larga 2, 3004-531 Coimbra, Portugal.
J Clin Med. 2023 Jan 20;12(3):855. doi: 10.3390/jcm12030855.
The major problems for the osseointegration of dental implants are the loosening of the screw that fixes the dental implant to the abutment and the micromovements that are generated when mechanical loads are applied. In this work, torque differences in the tightening and loosening of the connection screws after 1 cycle, 10 cycles and 1000 cycles for 4 dental implants with 2 external and 2 internal connections were analyzed. The loosening of 240 implants (60 for each system) was determined using high-precision torsimeters and an electromechanical testing machine. A total of 60 dental implants for each of the 4 systems were inserted into fresh bovine bone to determine the micromovements. The implant stability values (ISQ) were determined by RFA. The mechanical loads were performed at 30° from 20 N to 200 N. By means of the Q-star technique, the micromovements were determined. It was observed that, for a few cycles, the loosening of the screw did not exceed a loss of tightening of 10% for both connections. However, for 1000 cycles, the loss for the external connection was around 20% and for the internal connection it was 13%. The micromovements showed a lineal increase with the applied load for the implant systems studied. An external connection presented greater micromotions for each level of applied load and lower ISQ values than internal ones. An excellent lineal correlation between the ISQ and micromobility was observed. These results may be very useful for clinicians in the selection of the type of dental implant, depending on the masticatory load of the patient as well as the consequences of the insertion torque of the dental implant and its revisions.
牙种植体骨结合的主要问题是将牙种植体固定到基台的螺钉松动以及施加机械负荷时产生的微动。在这项研究中,分析了2种外连接和2种内连接的4种植体在1个循环、10个循环和1000个循环后连接螺钉拧紧和松开时的扭矩差异。使用高精度扭力计和机电测试机确定240颗种植体(每个系统60颗)的松动情况。将4个系统中每个系统的60颗牙种植体植入新鲜牛骨中以确定微动情况。通过共振频率分析(RFA)确定种植体稳定性值(ISQ)。在与水平方向成30°角的方向上施加20 N至200 N的机械负荷。借助Q-star技术确定微动情况。观察到,在几个循环中,两种连接方式下螺钉的松动拧紧损失均未超过10%。然而,在1000个循环时,外连接的损失约为20%,内连接的损失为13%。在所研究的种植体系统中,微动随施加负荷呈线性增加。对于每个施加负荷水平,外连接比内连接表现出更大的微动和更低的ISQ值。观察到ISQ与微移动性之间存在良好的线性相关性。这些结果对于临床医生根据患者的咀嚼负荷以及牙种植体植入扭矩及其修正的后果来选择牙种植体类型可能非常有用。