The purpose of this study was to compare central corneal thickness (CCT) values and evaluate the agreement obtained with three different devices in healthy eyes.
A total of 120 eyes of 60 healthy individuals (36 men and 24 women) were enrolled in this retrospective study. CCT measurements were performed using an optical biometer (AL-Scan), spectral-domain optical coherence tomography (SD-OCT) (Topcon 3D) and ultrasonic pachymetry (UP) (Accupach VI), and the results were compared. Bland-Altman analysis was used to quantify the agreement between methods.
The mean patient age was 28±5.73years (18-40years). The mean CCT values obtained by AL-Scan, UP, and SD-OCT were 532.4μm±29.7, 549μm±30.4, and 547μm±30.6, respectively. The mean differences in CCT were 15.30±9.52μm between AL-Scan and OCT (P<0.01), 17.15±8.42μm between AL-Scan and UP (P<0.01), and 1.85± 8.78μm between UP and OCT (P=0.067). All three methods of CCT measurement were closely correlated with each other.
The present study results suggest that, despite good agreement between the three devices, AL-Scan significantly underestimated CCT compared to UP and OCT. Therefore, clinicians should be aware that different results can be obtained using different devices for CCT measurements. It would be a better approach not to use them as interchangeable in clinical practice. CCT examination and follow-up should be performed using the same device, especially for patients who will undergo refractive surgery.
本研究旨在比较中央角膜厚度(CCT)值,并评估三种不同设备在健康眼中获得的一致性。
本回顾性研究共纳入 60 名健康个体(36 名男性和 24 名女性)的 120 只眼。使用光学生物测量仪(AL-Scan)、谱域光相干断层扫描(SD-OCT)(Topcon 3D)和超声角膜测厚仪(UP)(Accupach VI)进行 CCT 测量,并对结果进行比较。采用 Bland-Altman 分析量化各方法之间的一致性。
患者平均年龄为 28±5.73 岁(18-40 岁)。AL-Scan、UP 和 SD-OCT 测量的平均 CCT 值分别为 532.4μm±29.7、549μm±30.4 和 547μm±30.6。AL-Scan 与 OCT 之间 CCT 的平均差异为 15.30±9.52μm(P<0.01),AL-Scan 与 UP 之间 CCT 的平均差异为 17.15±8.42μm(P<0.01),UP 与 OCT 之间 CCT 的平均差异为 1.85±8.78μm(P=0.067)。三种 CCT 测量方法之间均具有良好的相关性。
本研究结果表明,尽管三种设备之间具有良好的一致性,但与 UP 和 OCT 相比,AL-Scan 显著低估了 CCT。因此,临床医生应注意到,使用不同设备进行 CCT 测量可能会得到不同的结果。在临床实践中,最好不要将它们视为可互换的。尤其是对于将要接受屈光手术的患者,应使用相同的设备进行 CCT 检查和随访。