三种不同成像设备测量中央角膜厚度的比较。

Comparison of central corneal thickness measurements using three different imaging devices.

机构信息

Medicana International Hospital, Department of Ophthalmology, Samsun, Turkey.

Medicana International Hospital, Department of Ophthalmology, Samsun, Turkey.

出版信息

J Fr Ophtalmol. 2023 Jun;46(6):589-595. doi: 10.1016/j.jfo.2022.09.029. Epub 2023 Apr 17.

Abstract

PURPOSE

The purpose of this study was to compare central corneal thickness (CCT) values and evaluate the agreement obtained with three different devices in healthy eyes.

METHODS

A total of 120 eyes of 60 healthy individuals (36 men and 24 women) were enrolled in this retrospective study. CCT measurements were performed using an optical biometer (AL-Scan), spectral-domain optical coherence tomography (SD-OCT) (Topcon 3D) and ultrasonic pachymetry (UP) (Accupach VI), and the results were compared. Bland-Altman analysis was used to quantify the agreement between methods.

MAIN RESULTS

The mean patient age was 28±5.73years (18-40years). The mean CCT values obtained by AL-Scan, UP, and SD-OCT were 532.4μm±29.7, 549μm±30.4, and 547μm±30.6, respectively. The mean differences in CCT were 15.30±9.52μm between AL-Scan and OCT (P<0.01), 17.15±8.42μm between AL-Scan and UP (P<0.01), and 1.85± 8.78μm between UP and OCT (P=0.067). All three methods of CCT measurement were closely correlated with each other.

CONCLUSION

The present study results suggest that, despite good agreement between the three devices, AL-Scan significantly underestimated CCT compared to UP and OCT. Therefore, clinicians should be aware that different results can be obtained using different devices for CCT measurements. It would be a better approach not to use them as interchangeable in clinical practice. CCT examination and follow-up should be performed using the same device, especially for patients who will undergo refractive surgery.

摘要

目的

本研究旨在比较中央角膜厚度(CCT)值,并评估三种不同设备在健康眼中获得的一致性。

方法

本回顾性研究共纳入 60 名健康个体(36 名男性和 24 名女性)的 120 只眼。使用光学生物测量仪(AL-Scan)、谱域光相干断层扫描(SD-OCT)(Topcon 3D)和超声角膜测厚仪(UP)(Accupach VI)进行 CCT 测量,并对结果进行比较。采用 Bland-Altman 分析量化各方法之间的一致性。

主要结果

患者平均年龄为 28±5.73 岁(18-40 岁)。AL-Scan、UP 和 SD-OCT 测量的平均 CCT 值分别为 532.4μm±29.7、549μm±30.4 和 547μm±30.6。AL-Scan 与 OCT 之间 CCT 的平均差异为 15.30±9.52μm(P<0.01),AL-Scan 与 UP 之间 CCT 的平均差异为 17.15±8.42μm(P<0.01),UP 与 OCT 之间 CCT 的平均差异为 1.85±8.78μm(P=0.067)。三种 CCT 测量方法之间均具有良好的相关性。

结论

本研究结果表明,尽管三种设备之间具有良好的一致性,但与 UP 和 OCT 相比,AL-Scan 显著低估了 CCT。因此,临床医生应注意到,使用不同设备进行 CCT 测量可能会得到不同的结果。在临床实践中,最好不要将它们视为可互换的。尤其是对于将要接受屈光手术的患者,应使用相同的设备进行 CCT 检查和随访。