Keklik Muzaffer
Erciyes University, Kayseri, Turkey.
Hematol Transfus Cell Ther. 2024 Oct-Dec;46(4):335-339. doi: 10.1016/j.htct.2023.02.004. Epub 2023 Mar 17.
Autologous peripheral blood stem cell (PBSC) transplantation has become a standard treatment option for certain hematological malignancies. The collection of PBSCs for transplantation is a well-established process and the effectiveness can vary depending on the cell separator. We aimed to compare the effectivity of two devices, the Spectra Optia and the Amicus for autologous PBSC collection. We also evaluated the effect of the peripheral white blood cell (WBC) count on the CD34+ collection efficiency (CE2).
We retrospectively evaluated 262 apheresis procedures performed in patients between 2015 and 2021 at the Apheresis Unit of our transplantation center. The PBSCs were collected by the Spectra Optia cell separator with continuous Mononuclear Collection (cMNC) (128 procedures) or the Amicus (MNC) (134 procedures). In addition to the apheresis parameters and product characteristics, we also evaluated the effect of the pre-apheresis peripheral WBC count on the CE2.
There was no significant difference in the CD34+ CE2 between the Spectra Optia and Amicus devices (median 65.06% and 68.24%, respectively, p = 0.070). In the Amicus group, the CE2 ratio was found to be statistically significantly higher in patients with a pre-apheresis peripheral WBC count of 15 × 10/L (median 81.70%, 68.06%, 61.35% and 58.13%, respectively, p < 0.001).
While both devices collected autologous PBSC effectively and safely, the Amicus provided a higher rate of CE2 at low pre-apheresis WBC counts. To our knowledge, this is the first study to evaluate the CE2 in autologous PBSC collection devices based on pre-apheresis WBC counts.
自体外周血干细胞(PBSC)移植已成为某些血液系统恶性肿瘤的标准治疗选择。用于移植的PBSC采集是一个成熟的过程,其有效性可能因细胞分离仪而异。我们旨在比较两种设备——Spectra Optia和Amicus用于自体PBSC采集的效果。我们还评估了外周白细胞(WBC)计数对CD34+采集效率(CE2)的影响。
我们回顾性评估了2015年至2021年在我们移植中心单采室对患者进行的262次单采程序。PBSC通过采用连续单核细胞采集(cMNC)的Spectra Optia细胞分离仪采集(128次程序)或Amicus(MNC)采集(134次程序)。除了单采参数和产品特征外,我们还评估了单采前外周WBC计数对CE2的影响。
Spectra Optia和Amicus设备之间的CD34+ CE2无显著差异(中位数分别为65.06%和68.24%,p = 0.070)。在Amicus组中,发现单采前外周WBC计数为15×10/L的患者的CE2比率在统计学上显著更高(中位数分别为81.70%、68.06%、61.35%和58.13%,p < 0.001)。
虽然两种设备都能有效且安全地采集自体PBSC,但在单采前WBC计数较低时,Amicus的CE2率更高。据我们所知,这是第一项基于单采前WBC计数评估自体PBSC采集设备中CE2的研究。